Optetrak Offset Tibial Tray and Screws, Size 4F/3T, Alpha, Ref 208-04-41, Sterile, RX.
Published November 29, 2017 · Recall initiated: September 19, 2017
Summary
FDA records medical device Class II recall Z-0138-2018, published November 29, 2017 and initiated September 19, 2017. Recalling firm: Exactech, Inc.; Status: Terminated; Reported quantity: 527 devices; Lot: All lot numbers. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.
Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
There is a design incompatibility between certain sizes of the offset spine stiffener screw and the constrained condylar tibial insert when used in conjunction with an offset tibial tray.
What to do
Voluntary: Firm initiated · Letter
Get an email when a new FDA recall like this is published
Free. We only write when there's a new recall that matches. Unsubscribe in one click.
Recall details
- Manufacturer
- Exactech, Inc.
- Distributor
- Distribution was made to AL, FL, IN, ME, MN, NJ, NY, OH, OK, VA, and Puerto Rico. Foreign distribution was made to Argentina, Australia, Brazil, China, Colombia, France, India, Italy, Luxembourg, The Netherlands, Spain, Switzerland, Tunisia, and United Kingdom.
- Lot
- All lot numbers
- Risk classification
- Class II
- Area (FDA)
- device
- Country
- United States
Related recalls
- FDAExactech Equinoxe, Reverse Shoulder Humeral Liners, 322-38-00 (145-DEG PE 38MM HUM LINER +0), REF: 322-38-03September 24, 2025
- FDAAcuMatch XLE Item Number / Product Description: 140-28-27 AM EHXL 0® LNR G 28MM 140-32-27 AM EHXL 0® LNR GFebruary 19, 2025
- FDAAcumatch L-Series BIPOLAR LINER, 22mm (BIPOLAR ENDOPROSTHESIS-Hip), Item Numbers: a) 100-22-19, SZ. L; b)July 10, 2024
- FDAOPTETRAK Inset Patella, Item Numbers: a) 200-05-23, 23MM; b) 200-05-26, 26MM; c) 200-05-29, 29MMJune 26, 2024
- FDAOptetrak Offset Tibial Tray and Screws, Size 3F/2T, Alpha, Ref 208-04-31, Sterile, RX.November 29, 2017
- FDAOptetrak Offset Tibial Tray and Screws, Size 4F/3T, Beta, Ref 208-04-42, Sterile, RX.November 29, 2017
Questions about this recall
Was Optetrak Offset Tibial Tray and Screws, Size 4F/3T, Alpha, Ref 208-04-41, Sterile, RX. recalled?
Yes. FDA published recall Z-0138-2018 on November 29, 2017. Reported hazard: There is a design incompatibility between certain sizes of the offset spine stiffener screw and the constrained condylar tibial insert when used in conjunction with an offset tibial tray.
What should I do if I have this product?
Voluntary: Firm initiated · Letter