Brand Name: IOMAX Product Name: IOMAX Cortical Module Model/Catalog Number: 190296-200 UDI-DI code: 00840067100653 Product Description: The IOMAX Cortical Module contains a 16-channel amplifier, a low current electrical stimulator, a TCS stimulator, a visual stimulator and an auditory stimulator. The low current stimulator has one (1) constant current output pair with a current range of 0.01-20 mA. The TCS stimulator has nine (9) independent constant current/constant voltage outputs, with current range of 50-1500 mA and voltage range of 50-1000 V. Component: Is the cortical component of the IOMAX system
Published October 21, 2026 · Recall initiated: August 18, 2026
Summary
FDA records medical device Class II recall Z-0141-2027, published October 21, 2026 and initiated August 18, 2026. Recalling firm: Cadwell Industries Inc; Status: Ongoing; Reported quantity: 23; Lot: Model/Catalog Number: 190296-200 UDI-DI code: 00840067100653 Lot Code: These devices are identified by the serial numbers listed below: 19029607AA0326007 19029607AA0226013 19029607AA0426028 19029607AA0226010 19029607AA0226007 19029607AA0326026 19029607AA0226015 19029607AA0326020 19029607AA0326018 19029607AA0326021 19029607AA0326022 19029607AA0226009 19029607AA0526022 19029607AA0326017 19029607AA0326027 19029607AA0326016 19029607AA0226016 19029607AA0326028 19029607AA0326024 19029607AA0326025 19029607AA0326023 19029607AA0326031 19029607AA0326006. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.
Status when published: Ongoing — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
Stimulator device contains a defective printed circuit board assembly (PCBA) that may cause the device to unintentionally deliver electrical stimulation to the patient, or allow an unintended electrical current to reach the patient.
What to do
Voluntary: Firm initiated · Letter
Get an email when a new FDA recall like this is published
Free. We only write when there's a new recall that matches. Unsubscribe in one click.
Recall details
- Manufacturer
- Cadwell Industries Inc
- Distributor
- Worldwide - US Nationwide distribution in the state of TN and the countries of Spain, Vietnam, Malaysia, Saudi Arabia, Thailand, Israel, Mexico, Turkey, Oman, Colombia, Brazil, Belgium, Italy, South Africa, and Switzerland.
- Lot
- Model/Catalog Number: 190296-200 UDI-DI code: 00840067100653 Lot Code: These devices are identified by the serial numbers listed below: 19029607AA0326007 19029607AA0226013 19029607AA0426028 19029607AA0226010 19029607AA0226007 19029607AA0326026 19029607AA0226015 19029607AA0326020 19029607AA0326018 19029607AA0326021 19029607AA0326022 19029607AA0226009 19029607AA0526022 19029607AA0326017 19029607AA0326027 19029607AA0326016 19029607AA0226016 19029607AA0326028 19029607AA0326024 19029607AA0326025 19029607AA0326023 19029607AA0326031 19029607AA0326006
- Risk classification
- Class II
- Area (FDA)
- device
- Country
- United States
Related recalls
- FDABD Phoenix Panel NMIC-306. Catalog Number: 449292.October 21, 2026
- FDANeuroStar Advanced Therapy System. Contains the NeuroStar HP Coil Assembly Part Number 81-01000-200. AffectsOctober 21, 2026
- FDABD Phoenix Panel NMIC-312. Catalog Number: 449082.October 21, 2026
- FDANaviFlex LW Pusher; Advx Panc. Stent 4/5F; Material Number (UPN): M00535030;October 21, 2026
- FDACascade IOMAX Cortical Module, REF: 190296-200July 24, 2024
- FDACascade IOMAX Cortical Module, REF: 190296-200May 29, 2024
Questions about this recall
Was Brand Name: IOMAX Product Name: IOMAX Cortical Module Model/Catalog Number: 190296-200 UDI-DI code: 00840067100653 Product Description: The IOMAX Cortical Module contains a 16-channel amplifier, a low current electrical stimulator, a TCS stimulator, a visual stimulator and an auditory stimulator. The low current stimulator has one (1) constant current output pair with a current range of 0.01-20 mA. The TCS stimulator has nine (9) independent constant current/constant voltage outputs, with current range of 50-1500 mA and voltage range of 50-1000 V. Component: Is the cortical component of the IOMAX system recalled?
Yes. FDA published recall Z-0141-2027 on October 21, 2026. Reported hazard: Stimulator device contains a defective printed circuit board assembly (PCBA) that may cause the device to unintentionally deliver electrical stimulation to the patient, or allow an unintended electrical current to reach the patient.
What should I do if I have this product?
Voluntary: Firm initiated · Letter