ACCU-CHEK Spirit Combo insulin infusion pump sold as part of Accu-Chek Combo System Product Usage: The ACCU-CHEK Combo System is meant only for single-patient use; the device may not be used for screening or diagnosis of diabetes or for multiple patient use. It is indicated for the treatment of insulin requiring diabetes and for the quantitative measurement of glucose in fresh capillary whole blood from the finger. The ACCU-CHEK Spirit Combo Insulin Pump is intended for the subcutaneous continuous delivery of insulin, at set and variable rates, for the management of diabetes mellitus in persons requiring insulin as prescribed by a physician.
Published November 5, 2014 · Recall initiated: September 25, 2014
Summary
FDA records medical device Class II recall Z-0146-2015, published November 5, 2014 and initiated September 25, 2014. Recalling firm: Roche Diagnostics Operations, Inc.; Status: Terminated; Reported quantity: 8116; Lot: Model/Catalog/Part Number: Accu-Chek Combo Kit mg US/English version, 05458544001 Accu-Chek Combo Kit mg US/Spanish version, 055004686001 Accu-Chek Spirit Combo Replacement Pump US/EN, 05391512001 Accu-Chek Spirit Combo Demo Pump- US, 0539175000. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.
Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
The insulin pump may lose time and date settings during a power interruption (such as a battery change) due to a faulty capacitor. If the capacitor fails, the time and date will return to default setting. Though the pump prompts customers to confirm the time and date, if the user does not recognize the time and date have returned to default, a shift of their basal rate time block could occur.
What to do
Voluntary: Firm initiated · Letter
Get an email when a new FDA recall like this is published
Free. We only write when there's a new recall that matches. Unsubscribe in one click.
Recall details
- Manufacturer
- Roche Diagnostics Operations, Inc.
- Distributor
- US Nationwide Distribution in the states of CA, CT, IA, NV, OH, IN, WI, GA, MO, NY, VA, LA, ME, TX, NC, ND, SD, NJ, CO, AL, SC, IL, KY, FL, OR, MN, MD, MS, AR, WV, WA, TN, CT, ID, KS, MI, PA, UT, MA, RI, NM, OK, DE, AZ, NE, MT, NV, WY, AK, and NH.
- Lot
- Model/Catalog/Part Number: Accu-Chek Combo Kit mg US/English version, 05458544001 Accu-Chek Combo Kit mg US/Spanish version, 055004686001 Accu-Chek Spirit Combo Replacement Pump US/EN, 05391512001 Accu-Chek Spirit Combo Demo Pump- US, 0539175000
- Risk classification
- Class II
- Area (FDA)
- device
- Country
- United States
Related recalls
- FDAcobas HCYS, Homocysteine Enzymatic Assay, Material Number REF 06542921190, for use on cobas c 701/702, inNovember 27, 2024
- FDACreatine Kinase (CK) used on cobas c 311/501/502 and COBAS INTEGRA 400 plus-In vitro test for the quantitativeOctober 2, 2024
- FDAAlbumin Gen.2, Catalog 05166861190October 21, 2020
- FDAElecsys Anti-HAV IgM assay-for the in vitro qualitative detection of IgM antibodies to hepatitis A virusDecember 18, 2019
- FDAParker Eclipse Probe Cover, Latex Free, REF 38-01, 10 Eclipse Probe Covers, Size: 2.37/1.75"W x 9.5"L (60/44November 5, 2014
- FDABlunt Hurd Cannula Blunt Tip 10 mm with 5 mm Reducer simplifies open laparoscopy.November 5, 2014
Questions about this recall
Was ACCU-CHEK Spirit Combo insulin infusion pump sold as part of Accu-Chek Combo System Product Usage: The ACCU-CHEK Combo System is meant only for single-patient use; the device may not be used for screening or diagnosis of diabetes or for multiple patient use. It is indicated for the treatment of insulin requiring diabetes and for the quantitative measurement of glucose in fresh capillary whole blood from the finger. The ACCU-CHEK Spirit Combo Insulin Pump is intended for the subcutaneous continuous delivery of insulin, at set and variable rates, for the management of diabetes mellitus in persons requiring insulin as prescribed by a physician. recalled?
Yes. FDA published recall Z-0146-2015 on November 5, 2014. Reported hazard: The insulin pump may lose time and date settings during a power interruption (such as a battery change) due to a faulty capacitor. If the capacitor fails, the time and date will return to default setting. Though the pump prompts customers to confirm the time and date, if the user does not recognize the time and date have returned to default, a shift of their basal rate time block could occur.
What should I do if I have this product?
Voluntary: Firm initiated · Letter