Fresenius Liberty¿ Cycler Part Number: RTLR180111 Product Usage: The Liberty cycler is a portable device designed to perform acute and chronic peritoneal dialysis.
Published November 13, 2013 · Recall initiated: October 14, 2013
Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
Fluid leaking may be observed inside the pump compartment of the Liberty Cycler and may result in dialysate becoming contaminated, potentially causing peritonitis
What to do
Voluntary: Firm initiated · Two or more of the following: Email, Fax, Letter, Press Release, Telephone, Visit
Recall details
- Manufacturer
- Fresenius Medical Care Holdings, Inc.
- Distributor
- USA Nationwide Distribution
- Lot
- All serial numbers
- Risk classification
- Class II
- Area (FDA)
- device
- Country
- United States