Freedom EVO-2 100 Base Unit, with Air LiHa Arms The Freedom EVO is an open automation platform product for general laboratory use. It is intended for routine laboratory tasks, such as general purpose pipetting and general purpose liquid handling and robotic processes.
Published November 5, 2014 · Recall initiated: September 26, 2014
Summary
FDA records medical device Class II recall Z-0149-2015, published November 5, 2014 and initiated September 26, 2014. Recalling firm: Tecan US, Inc.; Status: Terminated; Reported quantity: 2 units; Lot: Part Number: 10641100, SN: 1407006311. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.
Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
Distortion (noise interference) of the liquid level capacitance can result in air aspiration instead of liquid aspiration. This might cause a false signal (air detection) with possible dispense errors resulting in erroneous results.
What to do
Voluntary: Firm initiated · Telephone
Get an email when a new FDA recall like this is published
Free. We only write when there's a new recall that matches. Unsubscribe in one click.
Recall details
- Manufacturer
- Tecan US, Inc.
- Distributor
- US Distribution to IL.
- Lot
- Part Number: 10641100, SN: 1407006311.
- Risk classification
- Class II
- Area (FDA)
- device
- Country
- United States
Related recalls
- FDAInfinite F500, in vitro diagnostic. Product Usage: The Tecan Infinite F500 is a multifunctional microplateJuly 8, 2015
- FDAsiemens Dimension® Phenytoin Flex® Reagent Cartridge The PTN method used on the Dimension® clinical chemistryNovember 5, 2014
- FDAAPTUS Ulna Shortening 2.5 Product Usage: APTUS Ulna Shortening 2.5 Plates are indicated for fractures andNovember 5, 2014
- FDABlunt Hurd Cannula Blunt Tip 10 mm with 5 mm Reducer simplifies open laparoscopy.November 5, 2014
- FDAJoerns EasyCare Bed Product Usage: The AC-powered adjustable hospital bed is a device intended for medicalNovember 5, 2014
- FDATecan Freedom EVO with Touch Tools Suite Version 3.0 The intended use of the Touch Tools Suite is to runJune 26, 2013
Questions about this recall
Was Freedom EVO-2 100 Base Unit, with Air LiHa Arms The Freedom EVO is an open automation platform product for general laboratory use. It is intended for routine laboratory tasks, such as general purpose pipetting and general purpose liquid handling and robotic processes. recalled?
Yes. FDA published recall Z-0149-2015 on November 5, 2014. Reported hazard: Distortion (noise interference) of the liquid level capacitance can result in air aspiration instead of liquid aspiration. This might cause a false signal (air detection) with possible dispense errors resulting in erroneous results.
What should I do if I have this product?
Voluntary: Firm initiated · Telephone