Freedom EVO-2 100 Base Unit, with Air LiHa Arms The Freedom EVO is an open automation platform product for general laboratory use. It is intended for routine laboratory tasks, such as general purpose pipetting and general purpose liquid handling and robotic processes.

Published November 5, 2014 · Recall initiated: September 26, 2014

Summary

FDA records medical device Class II recall Z-0149-2015, published November 5, 2014 and initiated September 26, 2014. Recalling firm: Tecan US, Inc.; Status: Terminated; Reported quantity: 2 units; Lot: Part Number: 10641100, SN: 1407006311. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.

Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.

What's the hazard

Distortion (noise interference) of the liquid level capacitance can result in air aspiration instead of liquid aspiration. This might cause a false signal (air detection) with possible dispense errors resulting in erroneous results.

What to do

Voluntary: Firm initiated · Telephone

Get an email when a new FDA recall like this is published

Free. We only write when there's a new recall that matches. Unsubscribe in one click.

Recall details

Manufacturer
Tecan US, Inc.
Distributor
US Distribution to IL.
Lot
Part Number: 10641100, SN: 1407006311.
Risk classification
Class II
Area (FDA)
device
Country
United States

Related recalls

Questions about this recall

Was Freedom EVO-2 100 Base Unit, with Air LiHa Arms The Freedom EVO is an open automation platform product for general laboratory use. It is intended for routine laboratory tasks, such as general purpose pipetting and general purpose liquid handling and robotic processes. recalled?

Yes. FDA published recall Z-0149-2015 on November 5, 2014. Reported hazard: Distortion (noise interference) of the liquid level capacitance can result in air aspiration instead of liquid aspiration. This might cause a false signal (air detection) with possible dispense errors resulting in erroneous results.

What should I do if I have this product?

Voluntary: Firm initiated · Telephone