Adjustable Gastric Band, packaged alone in transparent PETG inner and outer blisters with Tyvek lids. Product Usage: Intended for use in surgical treatment of morbid obesity
Published November 13, 2013 · Recall initiated: September 30, 2013
Summary
FDA records medical device Class II recall Z-0150-2014, published November 13, 2013 and initiated September 30, 2013. Recalling firm: Ethicon Endo-Surgery Inc; Status: Terminated; Reported quantity: Product Code: 2200-X - 2,475 units; Lot: The following product code and lot numbers were affected by this recall: Product Codes: 2200-X Lot Numbers: ZLGBC5, ZLNBCW, ZLNBJC, ZLPBB9, ZMBBBB, ZMCBBC, ZMGBCJ, ZMHBBT, ZMHBCV, ZMLBD0, ZMMBCD, ZMPBBC, ZNCBC6, ZNGBB3, and ZNGBBF. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.
Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
During aging studies, the firm determined that the gastric belts were unable to pass the test requirements established for Force to Lock.
What to do
Voluntary: Firm initiated · Letter
Get an email when a new FDA recall like this is published
Free. We only write when there's a new recall that matches. Unsubscribe in one click.
Recall details
- Manufacturer
- Ethicon Endo-Surgery Inc
- Distributor
- Worldwide Distribution: USA (nationwide) including the states of: Alabama, Connecticut, Indiana, Michigan, North Carolina, South Carolina, Alaska, Delaware, Kansas, Minnesota, Ohio, Tennessee, Arizona, Florida, Kentucky, Mississipp,i Oklahoma, Texas, Arkansas, Georgia, Louisiana, Nebraska, Oregon, Virginia, California, Idaho, Maryland, Nevada, Pennsylvania, Washington, Colorado, Illinois, Massachusetts, New Jersey, Rhode Island, and West Virginia; and the countries of: Argentina, France, Malaysia, Spain, Australia, Georgia, Martinique, Sweden, Belgium, Germany, Mexico, Switzerland, Brazil, Greece, Netherland, Taiwan, Chile, Hungary, Panama, Tunisia, China, India, Philippines, United Arab Emirates, Czech Republic, Israel, Poland, United Kingdom, Egypt, Italy, Portugal, Venezuela, Estonia, Kenya, Russia, Finland, Kuwait, and Saudi Arabia.
- Lot
- The following product code and lot numbers were affected by this recall: Product Codes: 2200-X Lot Numbers: ZLGBC5, ZLNBCW, ZLNBJC, ZLPBB9, ZMBBBB, ZMCBBC, ZMGBCJ, ZMHBBT, ZMHBCV, ZMLBD0, ZMMBCD, ZMPBBC, ZNCBC6, ZNGBB3, and ZNGBBF.
- Risk classification
- Class II
- Area (FDA)
- device
- Country
- United States
Related recalls
- FDAECHELON LINEAR Cutters Reload 80 mm Blue, intended for transection, resection, and/or creation of anastomoses,October 22, 2025
- FDAENDOPATH ECHELON 45mm RELOADS WITH GRIPPING SURFACE TECHNOLOGY (GST), ECHELON Endopath Stapler, EndoscopicFebruary 7, 2024
- FDALigamax 5 mm Endoscopic Multiple Clip ApplierDecember 18, 2019
- FDAECHELON FLEX Powered Plus ENDOPATH 60mm Stapler - 280mm shaft, Product Code PCEE60ANovember 6, 2019
- FDAAdjustable Gastric Band, packaged alone in transparent PETG inner and outer blisters with Tyvek lids, or inNovember 13, 2013
- FDAAdjustable Gastric Band, packaged alone in transparent PETG inner and outer blisters with Tyvek lids. ProductNovember 13, 2013
Questions about this recall
Was Adjustable Gastric Band, packaged alone in transparent PETG inner and outer blisters with Tyvek lids. Product Usage: Intended for use in surgical treatment of morbid obesity recalled?
Yes. FDA published recall Z-0150-2014 on November 13, 2013. Reported hazard: During aging studies, the firm determined that the gastric belts were unable to pass the test requirements established for Force to Lock.
What should I do if I have this product?
Voluntary: Firm initiated · Letter