PERMA-HAND Silk Black Braided Sterile Nonabsorbable Surgical Suture, 1X18" TF-5, Product Code: N271H
Published October 21, 2026 · Recall initiated: September 8, 2026
Summary
FDA records medical device Class II recall Z-0153-2027, published October 21, 2026 and initiated September 8, 2026. Recalling firm: Ethicon, LLC; Status: Ongoing; Reported quantity: 1764; Lot: UDI-DI:Box/Unit: 30705031047369/10705031047365. Lot: RLBDPJ. Expiration: SEPTEMBER 30, 2026. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.
Status when published: Ongoing — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
Silk sutures, indicated for use in cardiovascular, ophthalmic and neurological procedures, packaged with alcohol-based fluid, not part of product design, that if discovered prior to use may cause surgical delay. If used, fluid could cause reduced suture tensile strength or knot security, tissue irritation/inflammatory tissue response, and treatment failure, including wound dehiscence/bleeding.
What to do
Voluntary: Firm initiated · Letter
Get an email when a new FDA recall like this is published
Free. We only write when there's a new recall that matches. Unsubscribe in one click.
Recall details
- Manufacturer
- Ethicon, LLC
- Distributor
- Worldwide - US Nationwide distribution in the states of SC, PA, TX, CA, ID, LA, MO, AL, NC, UT, IL, MA, OH, AZ, TN, OK, NE, ME, IN, WI, FL, CO, WY, NH, NY, NC, MN and the country of CA.
- Lot
- UDI-DI:Box/Unit: 30705031047369/10705031047365. Lot: RLBDPJ. Expiration: SEPTEMBER 30, 2026
- Risk classification
- Class II
- Area (FDA)
- device
- Country
- United States
Related recalls
- FDAETHILON Nylon Suture, Product Code 697GNovember 4, 2026
- FDALikorall Overhead Lifts Product Code: 3121001 Product Description: LIKORALL 200 Product Code: 3122005 ProductOctober 21, 2026
- FDARAPICIDE PA Part A High-Level Disinfectant, Model/Catalog Number: ML020115October 21, 2026
- FDAGeneric Name: BioFire® Joint Infection (JI) Panel Generic Name: BioFire® Joint Infection (JI) PanelOctober 21, 2026
- FDARAPICIDE PA Part A High-Level Disinfectant, Model/Catalog Number: ML020116October 21, 2026
- FDAGeneric Name: FilmArray® Pneumonia Panel plus Generic Name: FilmArray® Pneumonia Panel plus Model/CatalogOctober 21, 2026
Questions about this recall
Was PERMA-HAND Silk Black Braided Sterile Nonabsorbable Surgical Suture, 1X18" TF-5, Product Code: N271H recalled?
Yes. FDA published recall Z-0153-2027 on October 21, 2026. Reported hazard: Silk sutures, indicated for use in cardiovascular, ophthalmic and neurological procedures, packaged with alcohol-based fluid, not part of product design, that if discovered prior to use may cause surgical delay. If used, fluid could cause reduced suture tensile strength or knot security, tissue irritation/inflammatory tissue response, and treatment failure, including wound dehiscence/bleeding.
What should I do if I have this product?
Voluntary: Firm initiated · Letter