Gish 3/8 x 3/8" Straight Connector with Luer fitting, Item Numbers: EC2160B, EC2160S, Lot Numbers: 1210100010, 1215100026, 1217000138, Exp. Dates: 4/16/17, 5/31/15, 7/31/15. Product Usage: The connectors are indicated for use in connecting flexible tubing used in extracorporeal blood or fluid circuits. The connectors are distributed as sterile stand alones and bulk non-sterile.
Published October 31, 2012 · Recall initiated: September 14, 2012
Summary
FDA records medical device Class II recall Z-0157-2013, published October 31, 2012 and initiated September 14, 2012. Recalling firm: Sorin Group USA, Inc.; Status: Terminated; Reported quantity: 1,640 units; Lot: Item Numbers: EC2160B, EC2160S, Lot Numbers: 1210100010, 1215100026, 1217000138, Exp. Dates: 4/16/17, 5/31/15, 7/31/15. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.
Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
Due to a molding irregularity, certain lots of connectors have the potential to leak fluid or entrain air at the interface of the luer fitting and mating cap. Entrained air could result in patient injury.
What to do
Voluntary: Firm initiated · Letter
Get an email when a new FDA recall like this is published
Free. We only write when there's a new recall that matches. Unsubscribe in one click.
Recall details
- Manufacturer
- Sorin Group USA, Inc.
- Distributor
- US Nationwide Distribution in the states of: AL, AR, CA, CO, CT, FL, IL, MI, MN, MO, MS, NC, NY, OH, OK, OR, PA, SC, TN, TX, UT, VA and WI.
- Lot
- Item Numbers: EC2160B, EC2160S, Lot Numbers: 1210100010, 1215100026, 1217000138, Exp. Dates: 4/16/17, 5/31/15, 7/31/15.
- Risk classification
- Class II
- Area (FDA)
- device
- Country
- United States
Related recalls
- FDASorin Group Aortic Arch Cannula, 7 mm x 10 in x 3/8 in, Rx Only, Sterile Product Usage: The Aortic ArchAugust 16, 2017
- FDASORIN GROUP, SMART PERFUSION PACK, SMART 1/4" PEDS PACK, REF 627262502, STERILE EO, Rx Only, 1 EAAugust 9, 2017
- FDASORIN GROUP, SMART PERFUSION PACK, SMART CIRCULATORY ASSIST PACK, REF 627255003, STERILE EO, Rx Only, 1 EAAugust 9, 2017
- FDANON-STERILE SAMPLE, SORIN GROUP, SMART PERFUSION PACK, REF 627376401, 1 EA, NOT FOR CLINICAL USEAugust 9, 2017
- FDARepeater Pump, Class II device used for facilitating repeatable drug dosage distribution and reconstitution inOctober 31, 2012
- FDAGish 3/8 x 1/2" Reducer Connector with Luer fitting, Item Numbers: EC2180S, Lot Numbers: 1108900099,October 31, 2012
Questions about this recall
Was Gish 3/8 x 3/8" Straight Connector with Luer fitting, Item Numbers: EC2160B, EC2160S, Lot Numbers: 1210100010, 1215100026, 1217000138, Exp. Dates: 4/16/17, 5/31/15, 7/31/15. Product Usage: The connectors are indicated for use in connecting flexible tubing used in extracorporeal blood or fluid circuits. The connectors are distributed as sterile stand alones and bulk non-sterile. recalled?
Yes. FDA published recall Z-0157-2013 on October 31, 2012. Reported hazard: Due to a molding irregularity, certain lots of connectors have the potential to leak fluid or entrain air at the interface of the luer fitting and mating cap. Entrained air could result in patient injury.
What should I do if I have this product?
Voluntary: Firm initiated · Letter