FDAFDA · Food and drugs·Recall number Z-0162-2015

Merit Maestro Microcatheter, 2.4F 130cm Straight, REF28MC24130ST, Rx Only, Non-pyrogenic Sterile if package is unopened or undamaged.

Published November 12, 2014 · Recall initiated: September 26, 2014

Summary

FDA records medical device Class II recall Z-0162-2015, published November 12, 2014 and initiated September 26, 2014. Recalling firm: Merit Medical Systems, Inc.; Status: Terminated; Reported quantity: 56 units; Lot: Catalog No: 28MC24130ST, Lot H665694, H669824, Exp 2017 and Catalog No: 28MC24150SN, Lot H665696, Exp 7/31/2014. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.

Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.

What's the hazard

These microcatheters were packaged with a 2.8 F distal tip, rather than the labeled 2.4 F.

What to do

Voluntary: Firm initiated · Letter

Get an email when a new FDA recall like this is published

Free. We only write when there's a new recall that matches. Unsubscribe in one click.

Recall details

Manufacturer
Merit Medical Systems, Inc.
Distributor
Nationwide and to India. No govt/military/VA consignees.
Lot
Catalog No: 28MC24130ST, Lot H665694, H669824, Exp 2017 and Catalog No: 28MC24150SN, Lot H665696, Exp 7/31/2014
Risk classification
Class II
Area (FDA)
device
Country
United States

Related recalls

Questions about this recall

Was Merit Maestro Microcatheter, 2.4F 130cm Straight, REF28MC24130ST, Rx Only, Non-pyrogenic Sterile if package is unopened or undamaged. recalled?

Yes. FDA published recall Z-0162-2015 on November 12, 2014. Reported hazard: These microcatheters were packaged with a 2.8 F distal tip, rather than the labeled 2.4 F.

What should I do if I have this product?

Voluntary: Firm initiated · Letter