Back to search
FDAFDA · Food and drugs·Recall number Z-0163-2013

Aviator Drill Bit (10 mm, 12 mm, 14 mm, 16 mm, Manufactured by: Stryker Spine SAS Z.1 Marticot -33610- Cestas France +33 (0) 5.57.97.06.30 http://www.stryker.com Distributed in the USA by Stryker Spine 2 Pearl Ct., Allendale NJ 07401-1677 USA +1-201-750-8000 Drill bits are reusable instruments, sterilized prior to use, used in conjunction with an Aviator Drill Guide to provide a pathway for screw insertion. The aviator Anterior Cervical Plating (ACP) System is intended for use as an aid in cervical spinal fusion and is intended for unilateral fixation.

Published November 7, 2012 · Recall initiated: July 5, 2012

Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.

What's the hazard

Aviator drill bits are 10 mm longer than the Reflex Hybrid drill bits and are not interchangeable which could lead to over-drilling by as much as 10 mm.

What to do

Voluntary: Firm initiated · Letter

Recall details

Manufacturer
Stryker Spine
Distributor
Nationwide distribution.
Lot
E207315 Catalog number 48770610, 48770612, 48770616 All lots
Risk classification
Class II
Area (FDA)
device
Country
United States