Grafton DMB Matrix, Model Numbers: a) T42200, b) T42210; Demineralized Bone Matrix
Published November 1, 2023 · Recall initiated: September 26, 2023
Summary
FDA records medical device Class II recall Z-0163-2024, published November 1, 2023 and initiated September 26, 2023. Recalling firm: Medtronic Sofamor Danek USA Inc; Status: Ongoing; Reported quantity: 377 units; Lot: a) T42200: GTIN 00643169123052, Serial Numbers: A59254-013, A57871-018, A59117-013, A59254-018, A59264-016, A58526-014, A58526-015, A58526-016, A58864-018, A58082-016, A58983-011, A58983-012, A59320-013, A59320-016, A57043-011, A58831-018, A59307-012, A59320-019, A58831-019, A59117-020, A57104-018, A58408-015, A58408-016, A58952-019, A59307-011, A59264-014, A58526-011, A58526-012, A58408-011, A58995-025, A59140-014, A59140-015, A59165-012, A56808-011, A56808-012, A57078-011, A57078-012, A58077-013, A59117-021, A59264-015, A57871-014, A56808-019, A56808-020, A57078-013, A57078-014, A57078-015, A57078-016, A57078-017, A58649-011, A58649-012, A58649-013, A58649-014, A58649-015, A58649-016, A58649-017, A58983-018, A58983-019, A58983-020, A59117-011, A59117-015, A58983-017, A58995-026, A56808-013, A58408-012, A57104-011, A58077-014, A59165-013, A59165-014, A58864-012, A56808-016, A56808-017, A57043-019, A57104-012, A57104-013, A57104-014, A57104-015, A58701-011, A58701-012, A58701-013, A58701-014, A58701-015, A58701-016, A58864-011, A58952-011, A58952-012, A58864-013, A58864-014, A58864-015, A58526-017, A59243-014, A58831-011, A58077-015, A58077-016, A58077-017, A58605-011, A58526-013, A59254-017, A57871-015, A57871-016, A58082-011, A59165-020, A59264-017, A59264-018, A56808-014, A58831-012, A59243-013, A59243-016, A59243-017, A59254-014, A59254-016, A59320-014, A58605-012, A58605-013, A58605-014, A58605-015, A57043-015, A58983-013, A58649-018, A58864-016, A58864-017, A59140-013, A59140-017, A59140-019, A59140-020, A59254-015, A59320-020, A59320-021, A57043-012, A58831-013, A58831-014, A56808-015, A57104-016, A58983-014, A58983-015, A58983-016, A59117-014, A59117-016, A59117-017, A57871-013, A58082-012, A58952-015, A59140-018, A59264-013, A58831-015, A58831-016, A58831-017, A58859-011, A57043-016, A58831-020, A58859-012, A58859-013, A58859-014, A59264-019, A59264-020, A59320-011, A59320-012, A59243-015, A59307-013, A59320-017, A59320-018, A58082-015, A58859-015, A59307-014, A59307-015, A58082-019, A58082-020, A58952-020, A59254-011, A57043-013, A58408-014, A59117-018, A59117-019, A58841-015, A58841-016, A58841-017, A58841-018, A59165-015, A59165-016, A59165-017, A59165-018, A57043-017, A57043-018, A58082-013, A58408-013, A59140-011, A59140-012, A57043-014, A58952-013, A58952-014, A58077-011, A58082-014, A59165-019, A57871-011, A57871-012, A58082-017, A58082-018, A59264-011, A59264-012, A58077-012, A59165-011, A58859-016, A58859-017, A57104-017, A58859-018, A56808-018, A57871-017, A58408-017, A59140-016, A58841-014, A58952-016, A59117-012, A59243-018, A59320-015, A59254-012; b) T42210: GTIN 00643169123069, Serial Numbers: A58133-043, A59122-023, A58133-046, A58518-019, A58518-020, A57362-017, A57362-019, A58545-022, A58133-045, A57362-011, A58545-028, A59122-022, A57344-020, A59122-017, A57344-021, A58133-044, A59122-018, A57344-022, A57344-018, A57344-019, A58545-024, A58518-015, A58545-011, A58545-012, A58133-047, A58518-011, A58518-012, A58518-013, A58518-014, A58122-011, A58122-012, A58122-013, A58122-014, A58122-015, A58122-016, A58122-017, A58122-018, A57344-023, A58518-016, A57387-026, A57387-027, A58518-017, A58518-018, A58545-013, A58545-014, A57387-029, A58518-031, A57344-024, A57387-011, A57387-012, A57387-013, A57387-014, A57387-015, A59122-019, A59122-020, A59122-021, A59122-024, A58518-021, A58518-022, A58518-023, A58518-024, A58518-025, A59200-013, A57362-012, A57362-013, A59200-014, A59200-015, A58545-016, A58545-023, A57387-024, A58545-015, A59122-025, A59122-026, A59122-027, A59200-011, A59200-012, A57344-013, A57344-025, A58545-017, A58545-018, A57344-017, A58133-041, A58133-042, A57387-016, A57387-017, A59200-016, A59200-017, A59200-018, A59200-019, A59200-020, A59200-021, A59200-022, A57387-028, A58133-036, A58133-040, A58545-029, A57362-014, A57362-015, A57362-016, A57387-018, A57387-019, A57387-020, A57387-021, A57387-022, A57387-023, A58122-019, A58122-020, A58122-021, A58122-022, A58122-023, A58133-039. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.
Status when published: Ongoing — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
Specific serial numbers of GraftonTM DBM product due to the potential that a step during the inspection of the outer Tyvek pouch may not have been performed according to requirements. Based on this, batches may have been released with potential non-conformances in the outer Tyvek pouch.
What to do
Voluntary: Firm initiated · Letter
Get an email when a new FDA recall like this is published
Free. We only write when there's a new recall that matches. Unsubscribe in one click.
Recall details
- Manufacturer
- Medtronic Sofamor Danek USA Inc
- Distributor
- US, Colombia, S. Korea, New Zealand, India, Taiwan
- Lot
- a) T42200: GTIN 00643169123052, Serial Numbers: A59254-013, A57871-018, A59117-013, A59254-018, A59264-016, A58526-014, A58526-015, A58526-016, A58864-018, A58082-016, A58983-011, A58983-012, A59320-013, A59320-016, A57043-011, A58831-018, A59307-012, A59320-019, A58831-019, A59117-020, A57104-018, A58408-015, A58408-016, A58952-019, A59307-011, A59264-014, A58526-011, A58526-012, A58408-011, A58995-025, A59140-014, A59140-015, A59165-012, A56808-011, A56808-012, A57078-011, A57078-012, A58077-013, A59117-021, A59264-015, A57871-014, A56808-019, A56808-020, A57078-013, A57078-014, A57078-015, A57078-016, A57078-017, A58649-011, A58649-012, A58649-013, A58649-014, A58649-015, A58649-016, A58649-017, A58983-018, A58983-019, A58983-020, A59117-011, A59117-015, A58983-017, A58995-026, A56808-013, A58408-012, A57104-011, A58077-014, A59165-013, A59165-014, A58864-012, A56808-016, A56808-017, A57043-019, A57104-012, A57104-013, A57104-014, A57104-015, A58701-011, A58701-012, A58701-013, A58701-014, A58701-015, A58701-016, A58864-011, A58952-011, A58952-012, A58864-013, A58864-014, A58864-015, A58526-017, A59243-014, A58831-011, A58077-015, A58077-016, A58077-017, A58605-011, A58526-013, A59254-017, A57871-015, A57871-016, A58082-011, A59165-020, A59264-017, A59264-018, A56808-014, A58831-012, A59243-013, A59243-016, A59243-017, A59254-014, A59254-016, A59320-014, A58605-012, A58605-013, A58605-014, A58605-015, A57043-015, A58983-013, A58649-018, A58864-016, A58864-017, A59140-013, A59140-017, A59140-019, A59140-020, A59254-015, A59320-020, A59320-021, A57043-012, A58831-013, A58831-014, A56808-015, A57104-016, A58983-014, A58983-015, A58983-016, A59117-014, A59117-016, A59117-017, A57871-013, A58082-012, A58952-015, A59140-018, A59264-013, A58831-015, A58831-016, A58831-017, A58859-011, A57043-016, A58831-020, A58859-012, A58859-013, A58859-014, A59264-019, A59264-020, A59320-011, A59320-012, A59243-015, A59307-013, A59320-017, A59320-018, A58082-015, A58859-015, A59307-014, A59307-015, A58082-019, A58082-020, A58952-020, A59254-011, A57043-013, A58408-014, A59117-018, A59117-019, A58841-015, A58841-016, A58841-017, A58841-018, A59165-015, A59165-016, A59165-017, A59165-018, A57043-017, A57043-018, A58082-013, A58408-013, A59140-011, A59140-012, A57043-014, A58952-013, A58952-014, A58077-011, A58082-014, A59165-019, A57871-011, A57871-012, A58082-017, A58082-018, A59264-011, A59264-012, A58077-012, A59165-011, A58859-016, A58859-017, A57104-017, A58859-018, A56808-018, A57871-017, A58408-017, A59140-016, A58841-014, A58952-016, A59117-012, A59243-018, A59320-015, A59254-012; b) T42210: GTIN 00643169123069, Serial Numbers: A58133-043, A59122-023, A58133-046, A58518-019, A58518-020, A57362-017, A57362-019, A58545-022, A58133-045, A57362-011, A58545-028, A59122-022, A57344-020, A59122-017, A57344-021, A58133-044, A59122-018, A57344-022, A57344-018, A57344-019, A58545-024, A58518-015, A58545-011, A58545-012, A58133-047, A58518-011, A58518-012, A58518-013, A58518-014, A58122-011, A58122-012, A58122-013, A58122-014, A58122-015, A58122-016, A58122-017, A58122-018, A57344-023, A58518-016, A57387-026, A57387-027, A58518-017, A58518-018, A58545-013, A58545-014, A57387-029, A58518-031, A57344-024, A57387-011, A57387-012, A57387-013, A57387-014, A57387-015, A59122-019, A59122-020, A59122-021, A59122-024, A58518-021, A58518-022, A58518-023, A58518-024, A58518-025, A59200-013, A57362-012, A57362-013, A59200-014, A59200-015, A58545-016, A58545-023, A57387-024, A58545-015, A59122-025, A59122-026, A59122-027, A59200-011, A59200-012, A57344-013, A57344-025, A58545-017, A58545-018, A57344-017, A58133-041, A58133-042, A57387-016, A57387-017, A59200-016, A59200-017, A59200-018, A59200-019, A59200-020, A59200-021, A59200-022, A57387-028, A58133-036, A58133-040, A58545-029, A57362-014, A57362-015, A57362-016, A57387-018, A57387-019, A57387-020, A57387-021, A57387-022, A57387-023, A58122-019, A58122-020, A58122-021, A58122-022, A58122-023, A58133-039
- Risk classification
- Class II
- Area (FDA)
- device
- Country
- United States
Related recalls
- FDAGrafton" DBM and Grafton Plus" DBM contain demineralized human bone tissue combined with an inert additive toOctober 21, 2026
- FDAGrafton" DBM and Grafton Plus" DBM contain demineralized human bone tissue combined with an inert additive toOctober 21, 2026
- FDAMonoject 20 mL Syringe Luer-Lock Tip Soft PackNovember 1, 2023
- FDARobotics Pack, REF SJVO56BNovember 1, 2023
- FDAEndoskeleton" TCS Interbody System with Titan nanoLOCK" Surface Technology, NanoLock, 6 Deg, Interbody,October 20, 2021
- FDAAccurian RF Ablation, Cannula pk, Accurian, 150mm, 20G, 10mm, CS Product Number: AC0012 GTIN / UPN:July 29, 2020
Questions about this recall
Was Grafton DMB Matrix, Model Numbers: a) T42200, b) T42210; Demineralized Bone Matrix recalled?
Yes. FDA published recall Z-0163-2024 on November 1, 2023. Reported hazard: Specific serial numbers of GraftonTM DBM product due to the potential that a step during the inspection of the outer Tyvek pouch may not have been performed according to requirements. Based on this, batches may have been released with potential non-conformances in the outer Tyvek pouch.
What should I do if I have this product?
Voluntary: Firm initiated · Letter