Dimension Vista Protein 1 Control M, PROT 1 CON M, Siemens Material Number (SMN) 10445920, Product Code KC716; IVD

Published December 6, 2017 · Recall initiated: May 23, 2017

Summary

FDA records medical device Class II recall Z-0175-2018, published December 6, 2017 and initiated May 23, 2017. Recalling firm: Siemens Healthcare Diagnostics, Inc.; Status: Terminated; Reported quantity: N/A; Lot: Lot numbers: 6BQM01, 6BQM01A, 6BQM01B, 6BQM01C,6BQM01D, 6BQM01E (exp. date 2017-06-17); 6FQM01, 6FQM01A (exp. date 2017-11-03) --- Lot # (UDI): 6BQM01 (008427680063336JQL01C20180131), 6BQM01A (008427680063406BQM01A20170617), 6BQM01B (008427680063406BQM01B20170617), 6BQM01C (008427680063406BQM01C20170617), 6BQM01D( 008427680063406BQM01D20170617), 6BQM01E (008427680063406BQM01E20170617), 6FQM01(008427680063406FQM0120171103), 6FQM01A (008427680063406FQM01A20171103). Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.

Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.

What's the hazard

Complaint of high amount of ceruloplasmin patient results below the reference range (< 20 mg/dL [0.2 g/L]. PROT1 CAL and corresponding PROT1 CON L, M, and H are inaccurately assigned for ceruloplasmin (CER).

What to do

Voluntary: Firm initiated · Letter

Get an email when a new FDA recall like this is published

Free. We only write when there's a new recall that matches. Unsubscribe in one click.

Recall details

Manufacturer
Siemens Healthcare Diagnostics, Inc.
Distributor
Worldwide Distribution - US (nationwide) Internationally to Austria, Australia, Belgium, Canada, France, Germany, Great Britain, Italy, Japan, Netherlands, Portugal, Slovakia, Spain, Switzerland and Republic of Korea South
Lot
Lot numbers: 6BQM01, 6BQM01A, 6BQM01B, 6BQM01C,6BQM01D, 6BQM01E (exp. date 2017-06-17); 6FQM01, 6FQM01A (exp. date 2017-11-03) --- Lot # (UDI): 6BQM01 (008427680063336JQL01C20180131), 6BQM01A (008427680063406BQM01A20170617), 6BQM01B (008427680063406BQM01B20170617), 6BQM01C (008427680063406BQM01C20170617), 6BQM01D( 008427680063406BQM01D20170617), 6BQM01E (008427680063406BQM01E20170617), 6FQM01(008427680063406FQM0120171103), 6FQM01A (008427680063406FQM01A20171103).
Risk classification
Class II
Area (FDA)
device
Country
United States

Related recalls

Questions about this recall

Was Dimension Vista Protein 1 Control M, PROT 1 CON M, Siemens Material Number (SMN) 10445920, Product Code KC716; IVD recalled?

Yes. FDA published recall Z-0175-2018 on December 6, 2017. Reported hazard: Complaint of high amount of ceruloplasmin patient results below the reference range (< 20 mg/dL [0.2 g/L]. PROT1 CAL and corresponding PROT1 CON L, M, and H are inaccurately assigned for ceruloplasmin (CER).

What should I do if I have this product?

Voluntary: Firm initiated · Letter