Medline surgical convenience kits labeled as: 1) CV BASIN PACK-LF, Medline Product number/SKU DYNJ0538615K; 2) CV BASIN PACK-LF, Medline product number/SKU DYNJ0538615J; 3) CV BASIN PACK-LF, Medline product number/SKU DYNJ0538615L.
Published October 21, 2026 · Recall initiated: September 2, 2026
Summary
FDA records medical device Class II recall Z-0175-2027, published October 21, 2026 and initiated September 2, 2026. Recalling firm: Medline Industries, LP; Status: Ongoing; Reported quantity: 235 kits; Lot: Medline Product Number/SKU DYNJ0538615L; UDI/DI 10198459414152 (each) 40198459414153 (case), Lot number 26BBA206; Medline Product Number/SKU DYNJ0538615L; UDI/DI 10198459414152 (each) 40198459414153 (case), Lot number 25LBQ023; Medline Product Number/SKU DYNJ0538615L; UDI/DI 10198459414152 (each) 40198459414153 (case), Lot number 25KBR149; Medline Product Number/SKU DYNJ0538615K; UDI/DI 10198459136429 (each) 40198459136420 (case), Lot number 25EBF588; Medline Product Number/SKU DYNJ0538615K; UDI/DI 10198459136429 (each) 40198459136420 (case), Lot number 25ABI296; Medline Product Number/SKU DYNJ0538615K; UDI/DI 10198459136429 (each) 40198459136420 (case), Lot number 24IBP190; Medline Product Number/SKU DYNJ0538615J; UDI/DI 10195327673376 (each) 40195327673377 (case), Lot number 24DBC864. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.
Status when published: Ongoing — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
Medical convenience kits were distributed containing cannula components which do not conform to ISO 80369-7 dimensional requirements and may contribute to leaking or breaking during use.
What to do
Voluntary: Firm initiated · Two or more of the following: Email, Fax, Letter, Press Release, Telephone, Visit
Get an email when a new FDA recall like this is published
Free. We only write when there's a new recall that matches. Unsubscribe in one click.
Recall details
- Manufacturer
- Medline Industries, LP
- Distributor
- US Nationwide distribution.
- Lot
- Medline Product Number/SKU DYNJ0538615L; UDI/DI 10198459414152 (each) 40198459414153 (case), Lot number 26BBA206; Medline Product Number/SKU DYNJ0538615L; UDI/DI 10198459414152 (each) 40198459414153 (case), Lot number 25LBQ023; Medline Product Number/SKU DYNJ0538615L; UDI/DI 10198459414152 (each) 40198459414153 (case), Lot number 25KBR149; Medline Product Number/SKU DYNJ0538615K; UDI/DI 10198459136429 (each) 40198459136420 (case), Lot number 25EBF588; Medline Product Number/SKU DYNJ0538615K; UDI/DI 10198459136429 (each) 40198459136420 (case), Lot number 25ABI296; Medline Product Number/SKU DYNJ0538615K; UDI/DI 10198459136429 (each) 40198459136420 (case), Lot number 24IBP190; Medline Product Number/SKU DYNJ0538615J; UDI/DI 10195327673376 (each) 40195327673377 (case), Lot number 24DBC864.
- Risk classification
- Class II
- Area (FDA)
- device
- Country
- United States
Related recalls
- FDAMEDLINE medical convenience kits labeled as: 1) HIP SCOPE PACK, Medline Kit Number/SKU DYNJ64103C; 2) HIPOctober 28, 2026
- FDAMEDLINE medical convenience kits labeled as: 1) BIRTHDAY KIT-DOUBLE BASIN-LF, Medline Kit Number/SKUOctober 28, 2026
- FDAMEDLINE DRAPE C-SECTION FEN CLR SCREEN W, Medline Product Number/SKU DYNJP6108October 28, 2026
- FDAMEDLINE medical convenience kits labeled as: FETOSCOPY CDS-LF, Medline Product Number/SKU CDS840215POctober 28, 2026
- FDAMedline surgical convenience kits labeled as: 1) MAIN/LAPAROSCOPY TRANSPLA PACK, Medline product number/SKUOctober 21, 2026
- FDAMedline surgical convenience kits labeled as: 1) OPEN HEART PACK A & B, Medline Product number/SKUOctober 21, 2026
Questions about this recall
Was Medline surgical convenience kits labeled as: 1) CV BASIN PACK-LF, Medline Product number/SKU DYNJ0538615K; 2) CV BASIN PACK-LF, Medline product number/SKU DYNJ0538615J; 3) CV BASIN PACK-LF, Medline product number/SKU DYNJ0538615L. recalled?
Yes. FDA published recall Z-0175-2027 on October 21, 2026. Reported hazard: Medical convenience kits were distributed containing cannula components which do not conform to ISO 80369-7 dimensional requirements and may contribute to leaking or breaking during use.
What should I do if I have this product?
Voluntary: Firm initiated · Two or more of the following: Email, Fax, Letter, Press Release, Telephone, Visit