Back to search
FDAFDA · Food and drugs·Recall number Z-0188-2025

medex TranStar Disposable Transducer 1/EA, Product Code REF MX950

Published October 30, 2024 · Recall initiated: September 24, 2024

Summary

FDA records medical device Class II recall Z-0188-2025, published October 30, 2024 and initiated September 24, 2024. Recalling firm: Smiths Medical ASD, Inc.; Status: Ongoing; Reported quantity: 1100 units; Lot: UDI/DI 50351688504126 (case), 10351688504128 (pouch), Lot Numbers: 4293125, 4328466, 4277644, 4349995, 4363255. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.

Status when published: Ongoing — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.

What's the hazard

Due to a manufacturing defect, users may experience inaccurate pressure monitoring, an interruption in pressure monitoring, or an inability to zero the device due to either a leak within the TranStar Disposable Pressure Transducer or from a defect within the TranStar Disposable Pressure Transducer chip.

What to do

Voluntary: Firm initiated

Get an email when a new FDA recall like this is published

Free. We only write when there's a new recall that matches. Unsubscribe in one click.

Recall details

Manufacturer
Smiths Medical ASD, Inc.
Distributor
Worldwide distribution.
Lot
UDI/DI 50351688504126 (case), 10351688504128 (pouch), Lot Numbers: 4293125, 4328466, 4277644, 4349995, 4363255
Risk classification
Class II
Area (FDA)
device
Country
United States

Related recalls

Questions about this recall

Was medex TranStar Disposable Transducer 1/EA, Product Code REF MX950 recalled?

Yes. FDA published recall Z-0188-2025 on October 30, 2024. Reported hazard: Due to a manufacturing defect, users may experience inaccurate pressure monitoring, an interruption in pressure monitoring, or an inability to zero the device due to either a leak within the TranStar Disposable Pressure Transducer or from a defect within the TranStar Disposable Pressure Transducer chip.

What should I do if I have this product?

Voluntary: Firm initiated