SIGMA SPECTRUM Infusion Pump with Master Drug Library component (SHARP LCD screen, 55074-2 revision B) Product Usage: Intended to be used as a visual aid during the controlled administration of intravenous fluids. These fluids may include pharmaceutical drugs, blood, blood products and mixtures of required patient therapy.
Published November 20, 2013 · Recall initiated: September 24, 2013
Summary
FDA records medical device Class II recall Z-0189-2014, published November 20, 2013 and initiated September 24, 2013. Recalling firm: Baxter Healthcare Corp.; Status: Terminated; Reported quantity: 346 units; Lot: Product Code: 35700BAX & 35700ABB; Serial Numbers: 433713, 435563, 704056, 707290, 715315, 724216, 740127, 752318, 752447, 753675, 754947, 759766, 765934, 770034, 772017, 773470, 776507, 781016, 781969, 783352, 783454, 787567, 789720, 791225, 794828, 794990, 797185, 804798, 808393, 808764, 810971, 812076, 818718, 821213, 824067, 824198, 839143, 850028, 850577, 852007, 852431, 855450, 860201, 864799, 876765, 877548, 877944, 878637, 879810, 879886, 880022, 884707, 884778, 885565, 889416, 889814, 890949, 890987, 894957, 895413, 896662, 896912, 897576, 900401, 904778, 913511, 915035, 915981, 916173, 916948, 919159, 921162, 921937, 922515, 923861, 945583, 953011, 953360, 953995, 955881, 959847, 967694, 978892, 981692, 986022, 997333, 998512, 1002325, 1018432, 1018436, 1018737. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.
Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
Baxter Healthcare Corporation is recalling the LCD display screen on the SIGMA SPECTRUM Infusion Pump with Master Drug Library for not meeting standards for withstanding an electrostatic discharge.
What to do
Voluntary: Firm initiated · Letter
Get an email when a new FDA recall like this is published
Free. We only write when there's a new recall that matches. Unsubscribe in one click.
Recall details
- Manufacturer
- Baxter Healthcare Corp.
- Distributor
- Worldwide Distribution - USA Nationwide in the states of: AR, AZ, CA, CO, CT, FL, HI, IL, IN, KY, LA, MA, MD, MI, MN, MO, MS, NC, NH, NM, NY, OH, OK, OR, PA, TN, UT, VA, WA and WV and the country Canada.
- Lot
- Product Code: 35700BAX & 35700ABB; Serial Numbers: 433713, 435563, 704056, 707290, 715315, 724216, 740127, 752318, 752447, 753675, 754947, 759766, 765934, 770034, 772017, 773470, 776507, 781016, 781969, 783352, 783454, 787567, 789720, 791225, 794828, 794990, 797185, 804798, 808393, 808764, 810971, 812076, 818718, 821213, 824067, 824198, 839143, 850028, 850577, 852007, 852431, 855450, 860201, 864799, 876765, 877548, 877944, 878637, 879810, 879886, 880022, 884707, 884778, 885565, 889416, 889814, 890949, 890987, 894957, 895413, 896662, 896912, 897576, 900401, 904778, 913511, 915035, 915981, 916173, 916948, 919159, 921162, 921937, 922515, 923861, 945583, 953011, 953360, 953995, 955881, 959847, 967694, 978892, 981692, 986022, 997333, 998512, 1002325, 1018432, 1018436, 1018737
- Risk classification
- Class II
- Area (FDA)
- device
- Country
- United States
Related recalls
- FDAGAMBRO Capillary Dialyzer, Polyflux(R) Revaclear MAX, STERILE, Product code 110634 Product Usage: TheNovember 16, 2016
- FDASynovis VASCU-GUARD Peripheral Vascular Patch in the following sizes and product codes: 1 cm x 6 cm, productAugust 24, 2016
- FDA0.9% Sodium Chloride Injection, USP, 250 mL VIAFLEX Plastic Container Bag, Rx Only, Manufactured by BaxterJune 15, 2016
- FDADextrose Injection,USP,70% 2000 mL VIAFLEX Plastic Container Bag, Pharmacy Bulk Package, Not For DirectJune 15, 2016
- FDASpacelabs Healthcare elance Vital Signs Monitor, Model 93300, with Option S. It is a patient monitor withNovember 20, 2013
- FDASterrad 100NX, Product Code: 10104. Used to sterilize heat and moisture sensitive reusable medical devices.November 20, 2013
Questions about this recall
Was SIGMA SPECTRUM Infusion Pump with Master Drug Library component (SHARP LCD screen, 55074-2 revision B) Product Usage: Intended to be used as a visual aid during the controlled administration of intravenous fluids. These fluids may include pharmaceutical drugs, blood, blood products and mixtures of required patient therapy. recalled?
Yes. FDA published recall Z-0189-2014 on November 20, 2013. Reported hazard: Baxter Healthcare Corporation is recalling the LCD display screen on the SIGMA SPECTRUM Infusion Pump with Master Drug Library for not meeting standards for withstanding an electrostatic discharge.
What should I do if I have this product?
Voluntary: Firm initiated · Letter