FDAFDA · Food and drugs·Recall number Z-0190-2027

The ABL90 FLEX and ABL90 FLEX PLUS analyzers are intended for use by trained technologists, nurses, physicians and therapists. It is intended for use in a laboratory environment, near patient or point-of-care setting. These tests are only performed under a physician's order.

Published October 21, 2026 · Recall initiated: August 18, 2026

Summary

FDA records medical device Class II recall Z-0190-2027, published October 21, 2026 and initiated August 18, 2026. Recalling firm: Radiometer Medical ApS; Status: Ongoing; Reported quantity: 5395; Lot: UDI-DI: 05700693930923, 05700693930909 Software versions 3.6 and lower. ABL90 FLEX PLUS Instructions for Use versions prior to: 996-223, Version: 202606O. ABL90 FLEX Instructions for Use versions prior to: 996-189, Version: 202606L. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.

Status when published: Ongoing — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.

What's the hazard

In vitro diagnostic analyzer had a risk of potentially inaccurate results requiring changes to instructions for use and: 1) Sample types, including whole blood samples no longer suitable for testing FHbF and ctBil, 2) Reportable ranges for some parameters, 3) Analyzer configuration, and 4) limitations and warnings.

What to do

Voluntary: Firm initiated · Letter

Get an email when a new FDA recall like this is published

Free. We only write when there's a new recall that matches. Unsubscribe in one click.

Recall details

Manufacturer
Radiometer Medical ApS
Distributor
US Nationwide distribution including in the states of CA, OH, LA, MA, FL, NY, CO, IL, TX, GA, NC, PA, SC, MI, OK, WI, AR, MN, OR, UT, WA, SD, IN, TN, AZ, NE, WY, KY, AL, AK, ID, NJ, DE, MT, CT, MS, MO, VT, WV, IA, MD, KS, DC, NV, VA, NM, NH, ND, ME, HI, RI.
Lot
UDI-DI: 05700693930923, 05700693930909 Software versions 3.6 and lower. ABL90 FLEX PLUS Instructions for Use versions prior to: 996-223, Version: 202606O. ABL90 FLEX Instructions for Use versions prior to: 996-189, Version: 202606L
Risk classification
Class II
Area (FDA)
device
Country
Denmark

Related recalls

Questions about this recall

Was The ABL90 FLEX and ABL90 FLEX PLUS analyzers are intended for use by trained technologists, nurses, physicians and therapists. It is intended for use in a laboratory environment, near patient or point-of-care setting. These tests are only performed under a physician's order. recalled?

Yes. FDA published recall Z-0190-2027 on October 21, 2026. Reported hazard: In vitro diagnostic analyzer had a risk of potentially inaccurate results requiring changes to instructions for use and: 1) Sample types, including whole blood samples no longer suitable for testing FHbF and ctBil, 2) Reportable ranges for some parameters, 3) Analyzer configuration, and 4) limitations and warnings.

What should I do if I have this product?

Voluntary: Firm initiated · Letter