DePuy Mitek FMS Outflow Tubing with One-Way valve Product Code:284649 Product Usage: The Intermediary Tubing for FMS® Fluid Management Systems connects the Inflow Tubing to the arthroscopic sheath during arthroscopic procedures. The Polyvinyl Chloride (PVC) tubing integrates several functional components including: luer locks for attaching the tube to the Inflow tubing and sheath, clamps, and a one-way valve. The FMS Outflow Tubing with One-Way valve is sterile and intended for single use with the FMS Fluid Management Systems in a surgical setting by personnel trained in arthroscopy.
Published November 20, 2013 · Recall initiated: August 29, 2013
Summary
FDA records medical device Class II recall Z-0192-2014, published November 20, 2013 and initiated August 29, 2013. Recalling firm: DePuy Mitek, Inc., a Johnson & Johnson Co.; Status: Terminated; Reported quantity: 29280 units; Lot: All lot codes. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.
Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
Pillow valve included in the FMS Duo outflow tubing set and the FMS Solo intermediary tubing set may not perform as intended. may result in backflow of irrigation fluid into the One Day Set which may potentially lead to patient cross-contamination
What to do
Voluntary: Firm initiated · Two or more of the following: Email, Fax, Letter, Press Release, Telephone, Visit
Get an email when a new FDA recall like this is published
Free. We only write when there's a new recall that matches. Unsubscribe in one click.
Recall details
- Manufacturer
- DePuy Mitek, Inc., a Johnson & Johnson Co.
- Distributor
- Worldwide Distribution - USA (Nationwide) and in the countries of: Australia , Austria , Belgium , Canada , Czech Republic , Denmark , Finland , France, Germany , India, Ireland , Israel, Italy , Latvia , Netherlands , Norway , Nepal, Poland , Portugal , Spain , Sweden , Switzerland , United Kingdom.
- Lot
- All lot codes
- Risk classification
- Class II
- Area (FDA)
- device
- Country
- United States
Related recalls
- FDAHealix Knotless ADV BR 4.75 Suture AnchorDecember 18, 2019
- FDADePuy Mitek LATARJET EXPERIENCE Coracoid Drill Product Code: 288201 Product Usage: The Latarjet Cortical ScrewJune 28, 2017
- FDADePuy Mitek LATARJET EXPERIENCE Extraction Driver Product Code: 288216 Product Usage: The Latarjet CorticalJune 28, 2017
- FDADePuy Mitek LATARJET EXPERIENCE Top Hat Tap Product Code: 288203 Product Usage: The Latarjet Cortical ScrewJune 28, 2017
- FDADePuy Mitek FMS Intermediary Tubing with One-Way valve Product Code: 281142 Product Usage: The IntermediaryNovember 20, 2013
- FDASarns" Soft-Flow® Aortic Cannula without Suture Flange, angled tip, wire-reinforced with luer port, aorticNovember 20, 2013
Questions about this recall
Was DePuy Mitek FMS Outflow Tubing with One-Way valve Product Code:284649 Product Usage: The Intermediary Tubing for FMS® Fluid Management Systems connects the Inflow Tubing to the arthroscopic sheath during arthroscopic procedures. The Polyvinyl Chloride (PVC) tubing integrates several functional components including: luer locks for attaching the tube to the Inflow tubing and sheath, clamps, and a one-way valve. The FMS Outflow Tubing with One-Way valve is sterile and intended for single use with the FMS Fluid Management Systems in a surgical setting by personnel trained in arthroscopy. recalled?
Yes. FDA published recall Z-0192-2014 on November 20, 2013. Reported hazard: Pillow valve included in the FMS Duo outflow tubing set and the FMS Solo intermediary tubing set may not perform as intended. may result in backflow of irrigation fluid into the One Day Set which may potentially lead to patient cross-contamination
What should I do if I have this product?
Voluntary: Firm initiated · Two or more of the following: Email, Fax, Letter, Press Release, Telephone, Visit