Monarch Airway Clearance System, Model 1000, Product Numbers: PMACS1NA (has blue tooth capability only) and PMACS1WI (has WIFI capability). The firm name on the label is Hill-Rom Services Pte. Ltd., Singapore. The Monarch Airway Clearance System, Model 1000, is intended to provide Airway Clearance Therapy and promote bronchial drainage where external manipulation of the thorax is the physicians choice of treatment. It is indicated for patients having difficulty with secretion clearance or the presence of atelectasis caused by mucus plugging.

Published December 6, 2017 · Recall initiated: September 15, 2017

Summary

FDA records medical device Class II recall Z-0192-2018, published December 6, 2017 and initiated September 15, 2017. Recalling firm: HILL-ROM MANUFACTURING, INC.; Status: Terminated; Reported quantity: 60 devices; Lot: Serial numbers S082VM0092, S122VM0171, S128VM0205, S151VM0242, S082VM0093, S122VM0172, S132VM0210, S151VM0243, S082VM0095, S122VM0175, S132VM0212, S152VM0251, S095VM0119, S122VM0176, S132VM0213, S157VM0257, S095VM0124, S124VM0177, S135VM0216, S157VM0261, S095VM0125, S124VM0178, S136VM0219, S158VM0264, S114VM0152, S124VM0180, S136VM0222, S158VM0266, S119VM0158, S124VM0181, S136VM0223, S158VM0267, S119VM0160, S126VM0187, S138VM0224, S158VM0268, S119VM0161, S126VM0193, S138VM0226, S158VM0271, S119VM0162, S126VM0198, S146VM0227, S158VM0273, S122VM0165, S126VM0199, S146VM0229, S159VM0276, S122VM0166, S128VM0202, S146VM0230, S160VM0279, S122VM0168, S128VM0203, S151VM0238, S167VM0289, S122VM0170, S128VM0204, S151VM0240, and S167VM0291. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.

Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.

What's the hazard

Monarch Electromagnetic Interference (EMI) may disrupt the operation of implanted pacemakers or implantable cardioverter defibrillators as well as other devices that might be subject to an EMI if they come into direct contact with the Monarch device.

What to do

Voluntary: Firm initiated · Telephone

Get an email when a new FDA recall like this is published

Free. We only write when there's a new recall that matches. Unsubscribe in one click.

Recall details

Manufacturer
HILL-ROM MANUFACTURING, INC.
Distributor
USA (nationwide) Distribution to the states of: AZ, CA, CO, CT, FL, GA, IA, IL, IN, KY, MA, MI, MN, MO, NE, OH, OK, OR, TN, TX, VA, WA, WI, and WV. Internationally to Canada. There was no military/government distribution.
Lot
Serial numbers S082VM0092, S122VM0171, S128VM0205, S151VM0242, S082VM0093, S122VM0172, S132VM0210, S151VM0243, S082VM0095, S122VM0175, S132VM0212, S152VM0251, S095VM0119, S122VM0176, S132VM0213, S157VM0257, S095VM0124, S124VM0177, S135VM0216, S157VM0261, S095VM0125, S124VM0178, S136VM0219, S158VM0264, S114VM0152, S124VM0180, S136VM0222, S158VM0266, S119VM0158, S124VM0181, S136VM0223, S158VM0267, S119VM0160, S126VM0187, S138VM0224, S158VM0268, S119VM0161, S126VM0193, S138VM0226, S158VM0271, S119VM0162, S126VM0198, S146VM0227, S158VM0273, S122VM0165, S126VM0199, S146VM0229, S159VM0276, S122VM0166, S128VM0202, S146VM0230, S160VM0279, S122VM0168, S128VM0203, S151VM0238, S167VM0289, S122VM0170, S128VM0204, S151VM0240, and S167VM0291.
Risk classification
Class II
Area (FDA)
device
Country
United States

Related recalls

Questions about this recall

Was Monarch Airway Clearance System, Model 1000, Product Numbers: PMACS1NA (has blue tooth capability only) and PMACS1WI (has WIFI capability). The firm name on the label is Hill-Rom Services Pte. Ltd., Singapore. The Monarch Airway Clearance System, Model 1000, is intended to provide Airway Clearance Therapy and promote bronchial drainage where external manipulation of the thorax is the physicians choice of treatment. It is indicated for patients having difficulty with secretion clearance or the presence of atelectasis caused by mucus plugging. recalled?

Yes. FDA published recall Z-0192-2018 on December 6, 2017. Reported hazard: Monarch Electromagnetic Interference (EMI) may disrupt the operation of implanted pacemakers or implantable cardioverter defibrillators as well as other devices that might be subject to an EMI if they come into direct contact with the Monarch device.

What should I do if I have this product?

Voluntary: Firm initiated · Telephone