Back to search
FDAFDA · Food and drugs·Recall number Z-0198-2025

medex TranStar KIDS KIT 31IN. CLOSED BLOOD SAMPLE KIT with 30cc/hr Flush Device 10/EA, Product Code REF MX9544

Published October 30, 2024 · Recall initiated: September 24, 2024

Summary

FDA records medical device Class II recall Z-0198-2025, published October 30, 2024 and initiated September 24, 2024. Recalling firm: Smiths Medical ASD, Inc.; Status: Ongoing; Reported quantity: 900 units; Lot: UDI/DI 50351688503082 (case), 10351688503084 (pouch), Lot Numbers: 4226214, 4346067, 4350021, 4359831. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.

Status when published: Ongoing — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.

What's the hazard

Due to a manufacturing defect, users may experience inaccurate pressure monitoring, an interruption in pressure monitoring, or an inability to zero the device due to either a leak within the TranStar Disposable Pressure Transducer or from a defect within the TranStar Disposable Pressure Transducer chip.

What to do

Voluntary: Firm initiated

Get an email when a new FDA recall like this is published

Free. We only write when there's a new recall that matches. Unsubscribe in one click.

Recall details

Manufacturer
Smiths Medical ASD, Inc.
Distributor
Worldwide distribution.
Lot
UDI/DI 50351688503082 (case), 10351688503084 (pouch), Lot Numbers: 4226214, 4346067, 4350021, 4359831
Risk classification
Class II
Area (FDA)
device
Country
United States

Related recalls

Questions about this recall

Was medex TranStar KIDS KIT 31IN. CLOSED BLOOD SAMPLE KIT with 30cc/hr Flush Device 10/EA, Product Code REF MX9544 recalled?

Yes. FDA published recall Z-0198-2025 on October 30, 2024. Reported hazard: Due to a manufacturing defect, users may experience inaccurate pressure monitoring, an interruption in pressure monitoring, or an inability to zero the device due to either a leak within the TranStar Disposable Pressure Transducer or from a defect within the TranStar Disposable Pressure Transducer chip.

What should I do if I have this product?

Voluntary: Firm initiated