Back to search
FDAFDA · Food and drugs·Recall number Z-0198-2025

medex TranStar KIDS KIT 31IN. CLOSED BLOOD SAMPLE KIT with 30cc/hr Flush Device 10/EA, Product Code REF MX9544

Published October 30, 2024 · Recall initiated: September 24, 2024

Status when published: Ongoing — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.

What's the hazard

Due to a manufacturing defect, users may experience inaccurate pressure monitoring, an interruption in pressure monitoring, or an inability to zero the device due to either a leak within the TranStar Disposable Pressure Transducer or from a defect within the TranStar Disposable Pressure Transducer chip.

What to do

Voluntary: Firm initiated

Recall details

Manufacturer
Smiths Medical ASD, Inc.
Distributor
Worldwide distribution.
Lot
UDI/DI 50351688503082 (case), 10351688503084 (pouch), Lot Numbers: 4226214, 4346067, 4350021, 4359831
Risk classification
Class II
Area (FDA)
device
Country
United States

Sell products like this?

Recalert watches your whole catalog for recalls like this one, automatically.

Start free — no card