Brand Name: VACUETTE Product Name: LH Lithium Heparin Separator Model/Catalog Number: 454008P Product Description: Greiner VACUETTE Blood Collection Tube with Lithium Heparin and Gel Separator (13 x 75 mm, 3.5ml)
Published October 28, 2026 · Recall initiated: July 29, 2026
Summary
FDA records medical device Class II recall Z-0200-2027, published October 28, 2026 and initiated July 29, 2026. Recalling firm: Greiner Bio-One North America, Inc.; Status: Ongoing; Reported quantity: 860450; Lot: Lot Code: Item Number: 454008P Lot Number: B26043FM UDI: Rack: 29120017579032, Case: 39120017579039 Expiration Date: 2027-07-12. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.
Status when published: Ongoing — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
Lot B26043FM may include EDTA contamination which introduces potassium while chelating calcium in patient samples, leading to falsely elevated potassium (K+) and falsely decreased calcium (Ca++) results which have the potential to cause hyperkalemia.
What to do
Voluntary: Firm initiated · Letter
Get an email when a new FDA recall like this is published
Free. We only write when there's a new recall that matches. Unsubscribe in one click.
Recall details
- Manufacturer
- Greiner Bio-One North America, Inc.
- Distributor
- U.S. Nationwide distribution in the states of CO, FL, IL, MO, NE, NY, OH, PA, and VA.
- Lot
- Lot Code: Item Number: 454008P Lot Number: B26043FM UDI: Rack: 29120017579032, Case: 39120017579039 Expiration Date: 2027-07-12
- Risk classification
- Class II
- Area (FDA)
- device
- Country
- United States
Related recalls
- FDAPhilips Smart-hopping 2.0 AP 1.4 GHz; Product/Model Number: 867216;October 28, 2026
- FDAROTOR ASSY-SW40, Reference Number 331302, when the rotors are installed in the following centrifuges: OptimaOctober 28, 2026
- FDAVACUETTE® TUBE 8 ml CAT Serum Separator Clot Activator 16x100 red cap-yellow ring, non-ridgedJuly 24, 2024
- FDAVACUETTE TUBE 4 ml FX Sodium Fluoride/Potassium Oxalate, Item #454297, 13x75 grey cap-black ring, non-ridged.January 24, 2024
- FDAgreiner bio-one VACUETTE TUBE, 5 ml LH Lithium Heparin Sep, 13x100 green cap-yellow ring, non-ridged, 24 racksJanuary 26, 2022
- FDAVACUETTE Tube 2ml 9NC Coagulation Sodium Citrate 3.2%. Item no. 454322. Lot no. B1908355.April 22, 2020
Questions about this recall
Was Brand Name: VACUETTE Product Name: LH Lithium Heparin Separator Model/Catalog Number: 454008P Product Description: Greiner VACUETTE Blood Collection Tube with Lithium Heparin and Gel Separator (13 x 75 mm, 3.5ml) recalled?
Yes. FDA published recall Z-0200-2027 on October 28, 2026. Reported hazard: Lot B26043FM may include EDTA contamination which introduces potassium while chelating calcium in patient samples, leading to falsely elevated potassium (K+) and falsely decreased calcium (Ca++) results which have the potential to cause hyperkalemia.
What should I do if I have this product?
Voluntary: Firm initiated · Letter