medex TranStar KIDS KIT 10IN. CLOSED BLOOD SAMPLE KIT with 30cc/hr Flush Device 10/EA, Product Code REF MX9548

Published October 30, 2024 · Recall initiated: September 24, 2024

Status when published: Ongoing — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.

What's the hazard

Due to a manufacturing defect, users may experience inaccurate pressure monitoring, an interruption in pressure monitoring, or an inability to zero the device due to either a leak within the TranStar Disposable Pressure Transducer or from a defect within the TranStar Disposable Pressure Transducer chip.

What to do

Voluntary: Firm initiated

Recall details

Manufacturer
Smiths Medical ASD, Inc.
Distributor
Worldwide distribution.
Lot
UDI/DI 50351688503006 (case), 10351688503008 (pouch), Lot Numbers: 4221225, 4250996, 4297929, 4304897,4316671, 4323138, 4328494, 4387189, 4404713
Risk classification
Class II
Area (FDA)
device
Country
United States