Back to search
FDAFDA · Food and drugs·Recall number Z-0203-2025

medex TranStar SINGLE 84IN SET WITH ANTI-SHUNT STOPCOCK 10/EA, Product Code REF MXG300243R1

Published October 30, 2024 · Recall initiated: September 24, 2024

Summary

FDA records medical device Class II recall Z-0203-2025, published October 30, 2024 and initiated September 24, 2024. Recalling firm: Smiths Medical ASD, Inc.; Status: Ongoing; Reported quantity: 3080 units; Lot: UDI/DI 50351688518123 (case), 10351688518125 (pouch), Lot Numbers: 4282765, 4307302, 4309198, 4311461, 4316668, 4328487, 4350011, 4366881. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.

Status when published: Ongoing — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.

What's the hazard

Due to a manufacturing defect, users may experience inaccurate pressure monitoring, an interruption in pressure monitoring, or an inability to zero the device due to either a leak within the TranStar Disposable Pressure Transducer or from a defect within the TranStar Disposable Pressure Transducer chip.

What to do

Voluntary: Firm initiated

Get an email when a new FDA recall like this is published

Free. We only write when there's a new recall that matches. Unsubscribe in one click.

Recall details

Manufacturer
Smiths Medical ASD, Inc.
Distributor
Worldwide distribution.
Lot
UDI/DI 50351688518123 (case), 10351688518125 (pouch), Lot Numbers: 4282765, 4307302, 4309198, 4311461, 4316668, 4328487, 4350011, 4366881
Risk classification
Class II
Area (FDA)
device
Country
United States

Related recalls

Questions about this recall

Was medex TranStar SINGLE 84IN SET WITH ANTI-SHUNT STOPCOCK 10/EA, Product Code REF MXG300243R1 recalled?

Yes. FDA published recall Z-0203-2025 on October 30, 2024. Reported hazard: Due to a manufacturing defect, users may experience inaccurate pressure monitoring, an interruption in pressure monitoring, or an inability to zero the device due to either a leak within the TranStar Disposable Pressure Transducer or from a defect within the TranStar Disposable Pressure Transducer chip.

What should I do if I have this product?

Voluntary: Firm initiated