Back to search
FDAFDA · Food and drugs·Recall number Z-0207-2026

Salem Sump PVC Tubes: Product Code/Product Name: 264929 Salem Sump" Stomach Tube, Dual Lumen, 12 Fr/Ch (4.0 mm) 264945 Salem Sump" Stomach Tube, Dual Lumen, 14 Fr/Ch (4.7 mm) 264960 Salem Sump" Stomach Tube, Dual Lumen, 16 Fr/Ch (5.3 mm) 264986 Salem Sump" Stomach Tube, Dual Lumen, 18 Fr/Ch (6.0 mm) 5558264957 Salem Sump" COILED TUBE 16Fr/Ch (5.3 mm) 5558264965 Salem Sump" TUBE 16 Fr/Ch (5.3 mm) 5558264973 Salem Sump" COILED TUBE 18Fr/Ch (6.0 mm) 5558264981 Salem Sump" TUBE 18 Fr/Ch 18Fr/Ch (6.0 mm) 8888264911 Salem Sump" Dual Lumen Stomach Tube, 10 Fr/Ch (3.3 mm), 36" (91 cm) 8888268060 Salem Sump" Dual Lumen Stomach Tube 6 Fr/Ch (2.0 mm), 24" (61 cm) 8888268086 Salem Sump" Dual Lumen Stomach Tube, 8 Fr/Ch (2.7 mm), 24" (61 cm)

Published October 29, 2025 · Recall initiated: September 18, 2025

Summary

FDA records medical device Class II recall Z-0207-2026, published October 29, 2025 and initiated September 18, 2025. Recalling firm: Cardinal Health 200, LLC; Status: Ongoing; Reported quantity: all lots; Lot: Product Code/Product Name/UDI-DI: 264929 Salem Sump" Stomach Tube, Dual Lumen, 12 Fr/Ch (4.0 mm) 10884521067196 264945 Salem Sump" Stomach Tube, Dual Lumen, 14 Fr/Ch (4.7 mm) 10884521067202 264960 Salem Sump" Stomach Tube, Dual Lumen, 16 Fr/Ch (5.3 mm) 10884521067219 264986 Salem Sump" Stomach Tube, Dual Lumen, 18 Fr/Ch (6.0 mm) 10884521067226 5558264957 Salem Sump" COILED TUBE 16Fr/Ch (5.3 mm) 50192253008171 5558264965 Salem Sump" TUBE 16 Fr/Ch (5.3 mm) 50192253008195 5558264973 Salem Sump" COILED TUBE 18Fr/Ch (6.0 mm) 50192253008201 5558264981 Salem Sump" TUBE 18 Fr/Ch 18Fr/Ch (6.0 mm) 50192253008218 8888264911 Salem Sump" Dual Lumen Stomach Tube, 10 Fr/Ch (3.3 mm), 36" (91 cm) 10192253012453 8888268060 Salem Sump" Dual Lumen Stomach Tube 6 Fr/Ch (2.0 mm), 24" (61 cm) 10192253012972 8888268086 Salem Sump" Dual Lumen Stomach Tube, 8 Fr/Ch (2.7 mm), 24" (61 cm) 10192253012996 All Lots. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.

Status when published: Ongoing — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.

What's the hazard

Firm has received complaints regarding Anti-Reflux Valve (ARV) breakage due to excessive force during use. Firm is providing additional guidance for use. Expansion of RES 95214.

What to do

Voluntary: Firm initiated · Letter

Get an email when a new FDA recall like this is published

Free. We only write when there's a new recall that matches. Unsubscribe in one click.

Recall details

Manufacturer
Cardinal Health 200, LLC
Distributor
Worldwide - US Nationwide distribution including in the states of AK, AL, AR, AZ, CA, CO, CT, DC, DE, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, RI, SC, SD, TN, TX, UT, VA, VT, WA, WI, WV, WY, PR, GU and the countries of Australia, Austria, Bahrain, Belgium, Bulgaria, Canada, Chile, Colombia, Croatia, Czech Republic, Denmark, Ecuador, Estonia, Finland, France, French Polynesia, Germany, Greece, Hong Kong, Hungary, Ireland, Israel, Italy, Jordan, Africa, Kuwait, Luxembourg, Mexico Monaco, Netherlands, New Zealand, Northern Ireland, Norway, Pakistan, Panama, Philippines, Poland, Portugal, Qatar, Romania, Russia, Saudi Arabia, Singapore, Slovakia, Slovenia, South Africa, Spain, Sweden, Switzerland, Taiwan, Thailand, Taiwan, UAE, United Kingdom, Vietnam.
Lot
Product Code/Product Name/UDI-DI: 264929 Salem Sump" Stomach Tube, Dual Lumen, 12 Fr/Ch (4.0 mm) 10884521067196 264945 Salem Sump" Stomach Tube, Dual Lumen, 14 Fr/Ch (4.7 mm) 10884521067202 264960 Salem Sump" Stomach Tube, Dual Lumen, 16 Fr/Ch (5.3 mm) 10884521067219 264986 Salem Sump" Stomach Tube, Dual Lumen, 18 Fr/Ch (6.0 mm) 10884521067226 5558264957 Salem Sump" COILED TUBE 16Fr/Ch (5.3 mm) 50192253008171 5558264965 Salem Sump" TUBE 16 Fr/Ch (5.3 mm) 50192253008195 5558264973 Salem Sump" COILED TUBE 18Fr/Ch (6.0 mm) 50192253008201 5558264981 Salem Sump" TUBE 18 Fr/Ch 18Fr/Ch (6.0 mm) 50192253008218 8888264911 Salem Sump" Dual Lumen Stomach Tube, 10 Fr/Ch (3.3 mm), 36" (91 cm) 10192253012453 8888268060 Salem Sump" Dual Lumen Stomach Tube 6 Fr/Ch (2.0 mm), 24" (61 cm) 10192253012972 8888268086 Salem Sump" Dual Lumen Stomach Tube, 8 Fr/Ch (2.7 mm), 24" (61 cm) 10192253012996 All Lots
Risk classification
Class II
Area (FDA)
device
Country
United States

Related recalls

Questions about this recall

Was Salem Sump PVC Tubes: Product Code/Product Name: 264929 Salem Sump" Stomach Tube, Dual Lumen, 12 Fr/Ch (4.0 mm) 264945 Salem Sump" Stomach Tube, Dual Lumen, 14 Fr/Ch (4.7 mm) 264960 Salem Sump" Stomach Tube, Dual Lumen, 16 Fr/Ch (5.3 mm) 264986 Salem Sump" Stomach Tube, Dual Lumen, 18 Fr/Ch (6.0 mm) 5558264957 Salem Sump" COILED TUBE 16Fr/Ch (5.3 mm) 5558264965 Salem Sump" TUBE 16 Fr/Ch (5.3 mm) 5558264973 Salem Sump" COILED TUBE 18Fr/Ch (6.0 mm) 5558264981 Salem Sump" TUBE 18 Fr/Ch 18Fr/Ch (6.0 mm) 8888264911 Salem Sump" Dual Lumen Stomach Tube, 10 Fr/Ch (3.3 mm), 36" (91 cm) 8888268060 Salem Sump" Dual Lumen Stomach Tube 6 Fr/Ch (2.0 mm), 24" (61 cm) 8888268086 Salem Sump" Dual Lumen Stomach Tube, 8 Fr/Ch (2.7 mm), 24" (61 cm) recalled?

Yes. FDA published recall Z-0207-2026 on October 29, 2025. Reported hazard: Firm has received complaints regarding Anti-Reflux Valve (ARV) breakage due to excessive force during use. Firm is providing additional guidance for use. Expansion of RES 95214.

What should I do if I have this product?

Voluntary: Firm initiated · Letter