Techno-Path Multichem AE Unassayed Level 3, REF:AE623X
Published October 28, 2026 · Recall initiated: August 14, 2026
Summary
FDA records medical device Class II recall Z-0207-2027, published October 28, 2026 and initiated August 14, 2026. Recalling firm: TECHNO-PATH MANUFACTURING LTD.; Status: Ongoing; Reported quantity: 832 units; Lot: UDI:(01)05391523442824/-Lot # 65705263. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.
Status when published: Ongoing — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
A routine stability studies have identified a stability drift affecting the Ammonia (NH4+) analyte in specific lots of Multichem AE quality control material. Additional testing confirmed that the labelled shelf-life performance claim for the Ammonia analyte can no longer be supported for the affected lots. This issue is limited to the Ammonia analyte only. The impacted device may cause delay of issuing results to the physician. Delayed recognition and diagnosis of elevated levels may impact in timely treatment of life-threatening levels described within medical toxicity intervention decision points.
What to do
Voluntary: Firm initiated · E-Mail
Get an email when a new FDA recall like this is published
Free. We only write when there's a new recall that matches. Unsubscribe in one click.
Recall details
- Manufacturer
- TECHNO-PATH MANUFACTURING LTD.
- Distributor
- US Nationwide distribution in the state of Massachusetts.
- Lot
- UDI:(01)05391523442824/-Lot # 65705263
- Risk classification
- Class II
- Area (FDA)
- device
- Country
- Ireland
Related recalls
- FDATechno-Path Multichem AE Unassayed Level 1 REF:AE621XOctober 28, 2026
- FDATechno-Path Multichem AE Unassayed Level 2 REF: AE622XOctober 28, 2026
- FDAIon Torrent Genexus Dx with software, Catalog: A40003539, which supports the Ion Torrent Genexus Dx IntegratedOctober 28, 2026
- FDADirexion Torqueable Microcatheter, NGMC/SINGLE/021/STRAIGHT/1RO/105; Material Number: M001195200October 28, 2026
- FDAGeneric Name: Softmed Breathables Advanced Toric (fanfilcon A) Soft (Hydrophilic) Contact Lens; Model/CatalogOctober 28, 2026
- FDAMultichem ID-B. Model/Catalog Number: SR103B Intended for use as an unassayed positive qualitative qualityNovember 19, 2025
Questions about this recall
Was Techno-Path Multichem AE Unassayed Level 3, REF:AE623X recalled?
Yes. FDA published recall Z-0207-2027 on October 28, 2026. Reported hazard: A routine stability studies have identified a stability drift affecting the Ammonia (NH4+) analyte in specific lots of Multichem AE quality control material. Additional testing confirmed that the labelled shelf-life performance claim for the Ammonia analyte can no longer be supported for the affected lots. This issue is limited to the Ammonia analyte only. The impacted device may cause delay of issuing results to the physician. Delayed recognition and diagnosis of elevated levels may impact in timely treatment of life-threatening levels described within medical toxicity intervention decision points.
What should I do if I have this product?
Voluntary: Firm initiated · E-Mail