AD-TECH Spencer Probe Depth Electrode, REF SD04R-AP58X-000 4-Contact 1.12mm (A style) Depth Electrode

Published October 30, 2024 · Recall initiated: September 19, 2024

Summary

FDA records medical device Class II recall Z-0219-2025, published October 30, 2024 and initiated September 19, 2024. Recalling firm: Ad-Tech Medical Instrument Corporation; Status: Completed; Reported quantity: 2 units; Lot: UDI/DI 00841823107510, Lot Number 170052 858859860. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.

Status when published: Completed — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.

What's the hazard

Ad-Tech is recalling these A-Style SD Depth Electrodes because a label discrepancy was identified on A style contact SD electrodes where the MR Conditional symbol was incorrectly placed instead of the MR Unsafe symbol. The A style contacts were not cleared for MR Conditional use. Our records indicate you have received one or more of the affected lots listed above.

What to do

Voluntary: Firm initiated · E-Mail

Get an email when a new FDA recall like this is published

Free. We only write when there's a new recall that matches. Unsubscribe in one click.

Recall details

Manufacturer
Ad-Tech Medical Instrument Corporation
Distributor
US Nationwide distribution in the states of CA, OH.
Lot
UDI/DI 00841823107510, Lot Number 170052 858859860
Risk classification
Class II
Area (FDA)
device
Country
United States

Related recalls

Questions about this recall

Was AD-TECH Spencer Probe Depth Electrode, REF SD04R-AP58X-000 4-Contact 1.12mm (A style) Depth Electrode recalled?

Yes. FDA published recall Z-0219-2025 on October 30, 2024. Reported hazard: Ad-Tech is recalling these A-Style SD Depth Electrodes because a label discrepancy was identified on A style contact SD electrodes where the MR Conditional symbol was incorrectly placed instead of the MR Unsafe symbol. The A style contacts were not cleared for MR Conditional use. Our records indicate you have received one or more of the affected lots listed above.

What should I do if I have this product?

Voluntary: Firm initiated · E-Mail