Surgical Handpiece, Model: SGA-E Order Code: H083, Nakanishi Inc. The SGS/SGA Surgical Handpieces are intended for surgical use. Typical usage includes dental implant use: drilling and tapping the maxilla and mandible.
Published October 31, 2018 · Recall initiated: June 1, 2018
Summary
FDA records medical device Class II recall Z-0221-2019, published October 31, 2018 and initiated June 1, 2018. Recalling firm: Nakanishi Inc.; Status: Terminated; Reported quantity: 18; Lot: Previous Operation Manual: OM-H0035E001 Model Name: SGA-E, Order Code: H083 FL800010-FL800020; FLY00001-FLY00007. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.
Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
Reprocessing instructions for the SGS/SGA Surgical Handpieces have been updated with instructions of manual cleaning and specific parameters for autoclave sterilization . The previous operation manuals associated with the SGS/SGA products must be replaced by new operation manuals.
What to do
Voluntary: Firm initiated · Letter
Get an email when a new FDA recall like this is published
Free. We only write when there's a new recall that matches. Unsubscribe in one click.
Recall details
- Manufacturer
- Nakanishi Inc.
- Distributor
- U.S. Distribution to states of: CA, FL, GA, IL, WA.; and internationally to: Canada.
- Lot
- Previous Operation Manual: OM-H0035E001 Model Name: SGA-E, Order Code: H083 FL800010-FL800020; FLY00001-FLY00007.
- Risk classification
- Class II
- Area (FDA)
- device
- Country
- Japan
Related recalls
- FDASurgical Handpiece, Model: SGS-E, Order Code: H084, Nakanishi Inc. The SGS/SGA Surgical Handpieces areOctober 31, 2018
- FDABACT/ALERT VIRTUO system, B Unit, with Version R2.0 Firmware Product Usage: The BacT/ALERT VIRTU MicrobialOctober 31, 2018
- FDASTREAMLAB CORE UNIT / LYNX WITH IOMOctober 31, 2018
- FDA3.0mm Headless Compression Screw, Long Thread 18mm, Part Number 04.226.118 DePuy Synthes 3.0mm HeadlessOctober 31, 2018
- FDAAPTIO 15000 REFRIGERATED STORAGE MODULEOctober 31, 2018
- FDACHS Custom Convenience Kit-COTTON TIP APPLICATORS, 10'S (6") Product Number:R1112BOctober 31, 2018
Questions about this recall
Was Surgical Handpiece, Model: SGA-E Order Code: H083, Nakanishi Inc. The SGS/SGA Surgical Handpieces are intended for surgical use. Typical usage includes dental implant use: drilling and tapping the maxilla and mandible. recalled?
Yes. FDA published recall Z-0221-2019 on October 31, 2018. Reported hazard: Reprocessing instructions for the SGS/SGA Surgical Handpieces have been updated with instructions of manual cleaning and specific parameters for autoclave sterilization . The previous operation manuals associated with the SGS/SGA products must be replaced by new operation manuals.
What should I do if I have this product?
Voluntary: Firm initiated · Letter