83" (211 cm) Smallbore Ext Set w/Remv 8 Gang 4-Way NanoClave® Stopcock w/Baseplate, Rotating Luer;1 unit per pouch, 25 pouches per case. UDI:(01)10887709089508(17)250101(30)25(10)4548244. The Intravascular Administration Set is a single use, sterile, non-pyrogenic device which provides access for the administration of fluids from a container to a patient's vascular system through the administration set's needle or catheter (which is inserted into a vein).

Published October 28, 2020 · Recall initiated: July 6, 2020

Summary

FDA records medical device Class II recall Z-0221-2021, published October 28, 2020 and initiated July 6, 2020. Recalling firm: ICU Medical, Inc.; Status: Terminated; Reported quantity: Total of all products (Listed #1 thru 101) = 304735 units; Lot: Lot Number:4548244. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.

Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.

What's the hazard

Identification of a potential manufacturing defect on the internal surface of the NanoClave within specific lots of NanoClave sets, which may inhibit a proper seal with the NanoClave spike.

What to do

Voluntary: Firm initiated · Letter

Get an email when a new FDA recall like this is published

Free. We only write when there's a new recall that matches. Unsubscribe in one click.

Recall details

Manufacturer
ICU Medical, Inc.
Distributor
Worldwide Distribution: US (nationwide): AL, AZ, CA, CO, DE, FL,GA,ID, IL,IN, KY, LA, MA, MD, MI, MN, MO, MS, NC, NE, NH,NJ,NM,NV, NY, OH, OK, OR, PA,RI, TN, TX, UT, VA, WA, WI, and WV; and OUS countries of: Belgium, Brazil, Canada, France, Germany, Ireland, Japan, Martinique, Mexico, Poland, and Saudi Arabia.
Lot
Lot Number:4548244.
Risk classification
Class II
Area (FDA)
device
Country
United States

Related recalls

Questions about this recall

Was 83" (211 cm) Smallbore Ext Set w/Remv 8 Gang 4-Way NanoClave® Stopcock w/Baseplate, Rotating Luer;1 unit per pouch, 25 pouches per case. UDI:(01)10887709089508(17)250101(30)25(10)4548244. The Intravascular Administration Set is a single use, sterile, non-pyrogenic device which provides access for the administration of fluids from a container to a patient's vascular system through the administration set's needle or catheter (which is inserted into a vein). recalled?

Yes. FDA published recall Z-0221-2021 on October 28, 2020. Reported hazard: Identification of a potential manufacturing defect on the internal surface of the NanoClave within specific lots of NanoClave sets, which may inhibit a proper seal with the NanoClave spike.

What should I do if I have this product?

Voluntary: Firm initiated · Letter