VITROS Chemistry Products Lipase Slides, Product Codes 1668409 (contains 60 Slides per Cartridge) and 8297749 (contains 18 slides per cartridge)
Published November 15, 2023 · Recall initiated: September 12, 2023
Summary
FDA records medical device Class II recall Z-0236-2024, published November 15, 2023 and initiated September 12, 2023. Recalling firm: Ortho-Clinical Diagnostics, INc.; Status: Ongoing; Reported quantity: 342 units; Lot: Product Code 1668409: UDI-DI: 10758750009510, Affected Lots: 5905-3348-1557; Product Code 8297749: UDI-DI: 10758750004683, Affected Lots: 5905-3348-2899. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.
Status when published: Ongoing — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
Products 1668409 and 8297749 may contain VITROS Lipase Slides with an abnormality which could cause -46% to +86% bias in lipase results depending on how the slides are used. Falsely low levels could result in a missed or delayed diagnosis of acute pancreatitis and may lead to diagnostic confusion in the setting of acute abdominal symptoms. Falsely elevated levels would not result in patient harm but may cause diagnostic confusion and a delay in accurate diagnosis of a patient with abdominal pain.
What to do
Voluntary: Firm initiated · Letter
Get an email when a new FDA recall like this is published
Free. We only write when there's a new recall that matches. Unsubscribe in one click.
Recall details
- Manufacturer
- Ortho-Clinical Diagnostics, INc.
- Distributor
- US States: CA, CO, FL, GA, IL, IN, KS, NC, NY, TN, TX, VT. Canada, France, India, Japan, South Korea
- Lot
- Product Code 1668409: UDI-DI: 10758750009510, Affected Lots: 5905-3348-1557; Product Code 8297749: UDI-DI: 10758750004683, Affected Lots: 5905-3348-2899
- Risk classification
- Class II
- Area (FDA)
- device
- Country
- United States
Related recalls
- FDAArial Water Resistant Pendant with Long Wristband, Models: 59362 and 2560-59362; mobile patient alarmNovember 15, 2023
- FDAFogarty Thru-Lumen Embolectomy Catheter, REF: 12TLW805F35November 15, 2023
- FDAMini Fruit Jelly Cups; Mango flavor; UPC: 4 714379 321005, Net Wt. 52.9 ozNovember 15, 2023
- FDADisposable Spetzler-Malis Bipolar forceps, 18CM SM DISP BAYONET 0.5MM TIP, Catalog Number 6760-180-005November 15, 2023
- FDAMini Fruit Jelly Cups; Lychee; UPC: 4 714379 320992, Net Wt. 52.9 ozNovember 15, 2023
- FDADJO EMPOWR Knee Punch Handle, REF: 801-05-040November 15, 2023
Questions about this recall
Was VITROS Chemistry Products Lipase Slides, Product Codes 1668409 (contains 60 Slides per Cartridge) and 8297749 (contains 18 slides per cartridge) recalled?
Yes. FDA published recall Z-0236-2024 on November 15, 2023. Reported hazard: Products 1668409 and 8297749 may contain VITROS Lipase Slides with an abnormality which could cause -46% to +86% bias in lipase results depending on how the slides are used. Falsely low levels could result in a missed or delayed diagnosis of acute pancreatitis and may lead to diagnostic confusion in the setting of acute abdominal symptoms. Falsely elevated levels would not result in patient harm but may cause diagnostic confusion and a delay in accurate diagnosis of a patient with abdominal pain.
What should I do if I have this product?
Voluntary: Firm initiated · Letter