VITROS Chemistry Products Lipase Slides, Product Codes 1668409 (contains 60 Slides per Cartridge) and 8297749 (contains 18 slides per cartridge)
Published November 15, 2023 · Recall initiated: September 12, 2023
Status when published: Ongoing — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
Products 1668409 and 8297749 may contain VITROS Lipase Slides with an abnormality which could cause -46% to +86% bias in lipase results depending on how the slides are used. Falsely low levels could result in a missed or delayed diagnosis of acute pancreatitis and may lead to diagnostic confusion in the setting of acute abdominal symptoms. Falsely elevated levels would not result in patient harm but may cause diagnostic confusion and a delay in accurate diagnosis of a patient with abdominal pain.
What to do
Voluntary: Firm initiated · Letter
Recall details
- Manufacturer
- Ortho-Clinical Diagnostics, INc.
- Distributor
- US States: CA, CO, FL, GA, IL, IN, KS, NC, NY, TN, TX, VT. Canada, France, India, Japan, South Korea
- Lot
- Product Code 1668409: UDI-DI: 10758750009510, Affected Lots: 5905-3348-1557; Product Code 8297749: UDI-DI: 10758750004683, Affected Lots: 5905-3348-2899
- Risk classification
- Class II
- Area (FDA)
- device
- Country
- United States