Piranha Screw 4.0 x 12mm, used in Piranha Surgical Plate System The Piranha Surgical Plate System is intended to provide stabilization of the cervical vertebrae for various indications as part of the Omni Spine Anterior Cervical Plate System. The Omni Spine Anterior Cervical Plate System is intended for anterior screw fixation to the cervical spine (C2-C7) as an adjustment to fusion in the treatment of degenerative disc disease (DDD), sponylolisthesis, spinal stenosis, tumors, trauma (i.e. fracture).
Published December 20, 2017 · Recall initiated: November 13, 2017
Summary
FDA records medical device Class II recall Z-0239-2018, published December 20, 2017 and initiated November 13, 2017. Recalling firm: Amendia, Inc dba Spinal Elements; Status: Terminated; Reported quantity: 82 units; Lot: Part No. CSF-40-12, Lot No. 132635. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.
Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
Potential incorrect marking and color identification of Piranha screw. Some screws marked as 12 mm actually measure 14 mm.
What to do
Voluntary: Firm initiated · Letter
Get an email when a new FDA recall like this is published
Free. We only write when there's a new recall that matches. Unsubscribe in one click.
Recall details
- Manufacturer
- Amendia, Inc dba Spinal Elements
- Distributor
- US Distribution to the states of : AR, CA, GA, LA, MD, NV, NY, OH and TX.
- Lot
- Part No. CSF-40-12, Lot No. 132635
- Risk classification
- Class II
- Area (FDA)
- device
- Country
- United States
Related recalls
- FDAUniCel DxH600, Part Number B23858, automated hematology analyzer (Note: the DxH 600 is the same instrument asDecember 20, 2017
- FDAXcela PICC Kit with ENDEXO and PASV Valve Technology, XCELA PASV 5F DL 55CM MST-70 Nitinol KIT PG, UPNDecember 20, 2017
- FDAUniCel DxH800, Part Numbers: 629029 and B24802, automated hematology analyzerDecember 20, 2017
- FDAXCELA PASV 5F DL 55CM IR-145 Nitinol Wire KIT PG, UPN H965251260, Catalog No. 25-126 The Xcela with PASV,December 20, 2017
- FDARemisol Advance Software (Driver Architect) Catalog Number B32129 The Remisol Advance Data Manager (StandDecember 20, 2017
- FDAXCELA PASV 5F DL 55CM MST-70 KIT PG, UPN H965457320, Catalog No. 45-732 The Xcela with PASV, BioFlo PICC withDecember 20, 2017
Questions about this recall
Was Piranha Screw 4.0 x 12mm, used in Piranha Surgical Plate System The Piranha Surgical Plate System is intended to provide stabilization of the cervical vertebrae for various indications as part of the Omni Spine Anterior Cervical Plate System. The Omni Spine Anterior Cervical Plate System is intended for anterior screw fixation to the cervical spine (C2-C7) as an adjustment to fusion in the treatment of degenerative disc disease (DDD), sponylolisthesis, spinal stenosis, tumors, trauma (i.e. fracture). recalled?
Yes. FDA published recall Z-0239-2018 on December 20, 2017. Reported hazard: Potential incorrect marking and color identification of Piranha screw. Some screws marked as 12 mm actually measure 14 mm.
What should I do if I have this product?
Voluntary: Firm initiated · Letter