Back to search
FDAFDA · Food and drugs·Recall number Z-0250-2021

Clave Stopcock. 1 unit per pouch, 50 multivac pouches per case. UDI:(01)10887709039916(17)250201(30)50(10)4727496. The ICU Medical Clave Stopcock is a device used to administer or withdraw fluids that flow from a container to a patient's vascular system through a needle or catheter inserted into a vein or artery. This device is a four-way stopcock with an integral CLAVE side-port for closed, needleless access to the patient tubing or other fluid container.

Published October 28, 2020 · Recall initiated: July 6, 2020

Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.

What's the hazard

Identification of a potential manufacturing defect on the internal surface of the NanoClave within specific lots of NanoClave sets, which may inhibit a proper seal with the NanoClave spike.

What to do

Voluntary: Firm initiated · Letter

Recall details

Manufacturer
ICU Medical, Inc.
Distributor
Worldwide Distribution: US (nationwide): AL, AZ, CA, CO, DE, FL,GA,ID, IL,IN, KY, LA, MA, MD, MI, MN, MO, MS, NC, NE, NH,NJ,NM,NV, NY, OH, OK, OR, PA,RI, TN, TX, UT, VA, WA, WI, and WV; and OUS countries of: Belgium, Brazil, Canada, France, Germany, Ireland, Japan, Martinique, Mexico, Poland, and Saudi Arabia.
Lot
Lot Number:4727496.
Risk classification
Class II
Area (FDA)
device
Country
United States