Direxion Torqueable Microcatheter, NGMC/TRANSEND/021/BERN/1RO/155; Material Number: M001195850
Published October 28, 2026 · Recall initiated: September 16, 2026
Summary
FDA records medical device Class II recall Z-0261-2027, published October 28, 2026 and initiated September 16, 2026. Recalling firm: Boston Scientific Cork, Ltd.; Status: Ongoing; Reported quantity: 926; Lot: Material Number: M001195850; UDI-DI: 08714729839781; Batch Numbers: 37996606, 38022677, 38022678, 38065534, 38065535, 38112646, 38149837, 38224591, 38259492, 38290688, 38446231, 38769969, 38769970, 38810249, 38875417, 38936818, 38970730, 38970731, 38990886, 39074178, 40466463, 40466605, 40466606, 40470665, 40470666, 40476198, 40484940, 40485711, 40485712, 40485782, 40492387, 40507437, 40512910, 40512911, 40512912, 40512914, 40523982, 40524002, 40525713, 40531378, 40537147. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.
Status when published: Ongoing — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
The product is defective due to insufficient UV curing of the hydrophilic coating which is present on the outer surface of the distal segment of the microcatheters and raised ridges on nitinol sleeves due to laser cutting misalignment.
What to do
Voluntary: Firm initiated · Letter
Get an email when a new FDA recall like this is published
Free. We only write when there's a new recall that matches. Unsubscribe in one click.
Recall details
- Manufacturer
- Boston Scientific Cork, Ltd.
- Distributor
- Worldwide distribution US Nationwide and territory of Puerto Rico. The countries of Albania, Argentina, Australia, Austria, Belgium, Brazil, Canada, Chile, China, Colombia, Costa Rica, Croatia, Czech Republic, Denmark, Ecuador, Finland, France, Germany, Greece, Hungary, Ireland, Israel, Italy, Japan, Jordan, Kuwait, Lebanon, Lithuania, Mexico, Moldova, Netherlands, New Zealand, Norway, Panama, PERU, Poland, Portugal, Qatar, Romania, Russian Fed., Saudi Arabia, Serbia, Singapore, Slovakia, Slovenia, South Africa, Spain, Sweden, Switzerland, Tunisia, T®rkiye, UAE, United Kingdom.
- Lot
- Material Number: M001195850; UDI-DI: 08714729839781; Batch Numbers: 37996606, 38022677, 38022678, 38065534, 38065535, 38112646, 38149837, 38224591, 38259492, 38290688, 38446231, 38769969, 38769970, 38810249, 38875417, 38936818, 38970730, 38970731, 38990886, 39074178, 40466463, 40466605, 40466606, 40470665, 40470666, 40476198, 40484940, 40485711, 40485712, 40485782, 40492387, 40507437, 40512910, 40512911, 40512912, 40512914, 40523982, 40524002, 40525713, 40531378, 40537147;
- Risk classification
- Class II
- Area (FDA)
- device
- Country
- Ireland
Related recalls
- FDADirexion Torqueable Microcatheter, NGMC/SINGLE/027/STRAIGHT/1RO/155; Material Number: M001195420October 28, 2026
- FDADirexion Torqueable Microcatheter, NGMC/SINGLE/021/SWAN/1RO/155; Material Number: M001195310October 28, 2026
- FDADirexion Torqueable Microcatheter, NGMC/SINGLE/027/STRAIGHT/1RO/105; Material Number: M001195400October 28, 2026
- FDADirexion Torqueable Microcatheter, NGMC/TRANSEND/027/BERN/1RO/130; Material Number: M001195940October 28, 2026
- FDADirexion Torqueable Microcatheter, NGMC/TRANSEND/021/BERN/1RO/130; Material Number: M001195840October 28, 2026
- FDADirexion Torqueable Microcatheter, NGMC/TRANSEND/027/STRAIGHT/1RO/130; Material Number: M001195910October 28, 2026
Questions about this recall
Was Direxion Torqueable Microcatheter, NGMC/TRANSEND/021/BERN/1RO/155; Material Number: M001195850 recalled?
Yes. FDA published recall Z-0261-2027 on October 28, 2026. Reported hazard: The product is defective due to insufficient UV curing of the hydrophilic coating which is present on the outer surface of the distal segment of the microcatheters and raised ridges on nitinol sleeves due to laser cutting misalignment.
What should I do if I have this product?
Voluntary: Firm initiated · Letter