Quadra S Offset Broach Handle The offset broach handle is intended to attach to the broaches of different sizes as the surgeon broaches the femoral bone in preparation for the implantation of the femoral stem.
Published November 14, 2012 · Recall initiated: October 13, 2008
Summary
FDA records medical device Class II recall Z-0267-2013, published November 14, 2012 and initiated October 13, 2008. Recalling firm: Medacta Usa Inc; Status: Terminated; Reported quantity: 12 units; Lot: Ref: 01.10.10.122 (Lot # 085119), Ref: 01.10.10.123 (Lot # 085120), Ref: 01.10.10.124 (Lot # 085244), Ref: 01.10.10.125 (Lot # 085243). Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.
Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
Medacta USA Inc has initiated a recall on the Quadra S Offset Broach Handle because of the potential of the rivet that locks the metallic wire to the lever may become loose during a surgery and make it impossible to lock/unlock the broaches.
What to do
Voluntary: Firm initiated · Letter
Get an email when a new FDA recall like this is published
Free. We only write when there's a new recall that matches. Unsubscribe in one click.
Recall details
- Manufacturer
- Medacta Usa Inc
- Distributor
- Worldwide Distribution -- US including the states of Texas, Nevada and Colorado., and the country of Belgium.
- Lot
- Ref: 01.10.10.122 (Lot # 085119), Ref: 01.10.10.123 (Lot # 085120), Ref: 01.10.10.124 (Lot # 085244), Ref: 01.10.10.125 (Lot # 085243)
- Risk classification
- Class II
- Area (FDA)
- device
- Country
- United States
Related recalls
- FDAShort Monobloc Stem Broaches, Size/REF: 7/04.01.10.0970, 8/04.01.10.0971, 9/04.01.10.0972, 10/04.01.10.0973,July 8, 2026
- FDAMectaLIF ANTERIOR - Lag Plate Flush H10, REF 03.30.300, interbody fusion deviceApril 26, 2023
- FDAGMK Efficiency - GMK Sphere CR Insert Instrument Set Size 1June 15, 2022
- FDAThe GMK knee system: designed for cemented use in total knee arthroplasty. Sphere and Primary Inserts withDecember 27, 2017
- FDAHoffman LRF Telescopic Strut Short Stryker Stryker Trauma AG CH-2545 Seizach Distributed in the USA byNovember 14, 2012
- FDAThe brand name of the device is AlgiNot, a dental impression material. Product Part Numbers AlgiNot Intra KitNovember 14, 2012
Questions about this recall
Was Quadra S Offset Broach Handle The offset broach handle is intended to attach to the broaches of different sizes as the surgeon broaches the femoral bone in preparation for the implantation of the femoral stem. recalled?
Yes. FDA published recall Z-0267-2013 on November 14, 2012. Reported hazard: Medacta USA Inc has initiated a recall on the Quadra S Offset Broach Handle because of the potential of the rivet that locks the metallic wire to the lever may become loose during a surgery and make it impossible to lock/unlock the broaches.
What should I do if I have this product?
Voluntary: Firm initiated · Letter