Philips brand IntelliSpace PACS 4.4, Image Management System; The system is a software package used with general purpose computing hardware to acquire, store, distribute, process and display images and associated data throughout a clinical environment. Product is manufactured and distributed by Philips Healthcare Informatics, Inc. Foster City, CA IntelliSpace PACS are image management systems intended to be used by trained professionals, including but not limited to physicians, nurses and medical technicians. The system is a software package used with general purpose computing hardware to acquire, store, distribute, process and display images and associated data throughout a clinical environment. The software performs digital image processing, measurement, communication and storage. iSite supports receiving, sending, printing, storing and displaying studies received from the following modality types via DICOM: CT, MR, NM, US, XA, PET, DX, DR, RE, RT, MG, SC, VL, as well as hospital/radiology information systems.
Published November 20, 2013 · Recall initiated: October 14, 2013
Summary
FDA records medical device Class II recall Z-0269-2014, published November 20, 2013 and initiated October 14, 2013. Recalling firm: Philips Healthcare Informatics, Inc.; Status: Terminated; Reported quantity: 71 units; Lot: All units of these IntelliSpace PACS. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.
Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
Philips Healthcare became aware of a software problem with the IntelliSpace PACS system, in that a timing issue has been identified that may result in a condition where the device does not store additional images received before processing is complete. The missing images may contain pathology which may result in a different diagnosis if they had been available at time of interpretation.
What to do
Voluntary: Firm initiated · Letter
Get an email when a new FDA recall like this is published
Free. We only write when there's a new recall that matches. Unsubscribe in one click.
Recall details
- Manufacturer
- Philips Healthcare Informatics, Inc.
- Distributor
- Worldwide Distribution - USA (nationwide) and country: Canada.
- Lot
- All units of these IntelliSpace PACS.
- Risk classification
- Class II
- Area (FDA)
- device
- Country
- United States
Related recalls
- FDAPhilips IntelliSpace PACS 4.4 image management systems performs digital image processing, measurement andAugust 27, 2014
- FDAPhilips XRE, the dictation editor system included in the Philips XIRIS 8.1.50 XIRIS is a radiology informationMay 28, 2014
- FDACarestream Vue PACS; UPG MX-RAD SVR 75K EX/YR UNLIMITED The CARESTREAM PACS is an image management systemNovember 20, 2013
- FDAAW VolumeShare 4 (ADVANTAGE WORKSTATION 4.3.), AW VolumeShare 5 (AW VOLUMESHARE 5 WITH ANGIOVIZ OPTION), AWNovember 20, 2013
- FDACarestream Vue PACS; UPG MX-RAD SVR 50K EX/YR UNLIMITED The CARESTREAM PACS is an image management systemNovember 20, 2013
- FDASiemens syngo.plaza image processing system image processing radiological systemNovember 20, 2013
Questions about this recall
Was Philips brand IntelliSpace PACS 4.4, Image Management System; The system is a software package used with general purpose computing hardware to acquire, store, distribute, process and display images and associated data throughout a clinical environment. Product is manufactured and distributed by Philips Healthcare Informatics, Inc. Foster City, CA IntelliSpace PACS are image management systems intended to be used by trained professionals, including but not limited to physicians, nurses and medical technicians. The system is a software package used with general purpose computing hardware to acquire, store, distribute, process and display images and associated data throughout a clinical environment. The software performs digital image processing, measurement, communication and storage. iSite supports receiving, sending, printing, storing and displaying studies received from the following modality types via DICOM: CT, MR, NM, US, XA, PET, DX, DR, RE, RT, MG, SC, VL, as well as hospital/radiology information systems. recalled?
Yes. FDA published recall Z-0269-2014 on November 20, 2013. Reported hazard: Philips Healthcare became aware of a software problem with the IntelliSpace PACS system, in that a timing issue has been identified that may result in a condition where the device does not store additional images received before processing is complete. The missing images may contain pathology which may result in a different diagnosis if they had been available at time of interpretation.
What should I do if I have this product?
Voluntary: Firm initiated · Letter