S5 Mast Roller Pump 85, Item Number: 10-88-60 The Stockert S5/Sorin C5 System is intended to be used during cardiopulmonary bypass for procedures lasting six hours or less.
Published November 14, 2012 · Recall initiated: September 19, 2012
Summary
FDA records medical device Class II recall Z-0273-2013, published November 14, 2012 and initiated September 19, 2012. Recalling firm: Sorin Group USA, Inc.; Status: Terminated; Reported quantity: 3 units; Lot: Serial Numbers: 50E50766--50E50769, 50E50772-50E50774, 50E50783-50E50784. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.
Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
Sorin Group USA, Inc. has issued an Urgent Field Safety Notice regarding certain serial numbers of the Sorin Group S5 Perfusion System due to reports of a "Fault in motor controller" alarm message displayed on the system panel. This could lead to unintended pump stoppage which may result in permanent impairment or life-threatening injury. An out of specification electrical component was determin
What to do
Voluntary: Firm initiated · Letter
Get an email when a new FDA recall like this is published
Free. We only write when there's a new recall that matches. Unsubscribe in one click.
Recall details
- Manufacturer
- Sorin Group USA, Inc.
- Distributor
- Worldwide distribution: USA (nationwide) including states of: AZ, GA, MI, MS, NC, NY, PA, SC, and TX; and countries of: Austria, Australia, Belgium, Canada, People's Republic of China, Columbia, Denmark, Finland, Germany, India, Iran-Islamic Republic of, Ireland, Japan, Lebanon, Morocco, Pakistan, Poland, Russian Federation, Saudia Arabia, Spain, South Africa, Thailand, Turkey and United Kingdom.
- Lot
- Serial Numbers: 50E50766--50E50769, 50E50772-50E50774, 50E50783-50E50784.
- Risk classification
- Class II
- Area (FDA)
- device
- Country
- United States
Related recalls
- FDASorin Group Aortic Arch Cannula, 7 mm x 10 in x 3/8 in, Rx Only, Sterile Product Usage: The Aortic ArchAugust 16, 2017
- FDASORIN GROUP, SMART PERFUSION PACK, SMART 1/4" PEDS PACK, REF 627262502, STERILE EO, Rx Only, 1 EAAugust 9, 2017
- FDASORIN GROUP, SMART PERFUSION PACK, SMART CIRCULATORY ASSIST PACK, REF 627255003, STERILE EO, Rx Only, 1 EAAugust 9, 2017
- FDANON-STERILE SAMPLE, SORIN GROUP, SMART PERFUSION PACK, REF 627376401, 1 EA, NOT FOR CLINICAL USEAugust 9, 2017
- FDAS5 Double Roller Pump 85, Item Number: 10-85-00 The Stockert S5/Sorin C5 System is intended to be used duringNovember 14, 2012
- FDAS5 Mast Roller Pump 150, Item Number: 10-88-00 The Stockert S5/Sorin C5 System is intended to be used duringNovember 14, 2012
Questions about this recall
Was S5 Mast Roller Pump 85, Item Number: 10-88-60 The Stockert S5/Sorin C5 System is intended to be used during cardiopulmonary bypass for procedures lasting six hours or less. recalled?
Yes. FDA published recall Z-0273-2013 on November 14, 2012. Reported hazard: Sorin Group USA, Inc. has issued an Urgent Field Safety Notice regarding certain serial numbers of the Sorin Group S5 Perfusion System due to reports of a "Fault in motor controller" alarm message displayed on the system panel. This could lead to unintended pump stoppage which may result in permanent impairment or life-threatening injury. An out of specification electrical component was determin
What should I do if I have this product?
Voluntary: Firm initiated · Letter