Medtronic Activa PC, model 37601, and Activa RC, model 37612, Activa SC, models 37602 and 37603. Multi-program Rechargeable Neurostimulator for Deep Brain Stimulation.
Published November 20, 2013 · Recall initiated: September 30, 2013
Summary
FDA records medical device Class II recall Z-0279-2014, published November 20, 2013 and initiated September 30, 2013. Recalling firm: Medtronic Neuromodulation; Status: Terminated; Reported quantity: 62,634; Lot: all serial numbers. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.
Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
Medtronic notified customers regarding Loss of Stimulation of specific Implantable Neuro-Stimulators under the following specific conditions: Switching between groups with Multiple programs or creating program groups. Therapy can be restored with the Physician Recharge mode (RPM) of the Implantable Neurostimulator Recharger (INSR).
What to do
Voluntary: Firm initiated · Letter
Get an email when a new FDA recall like this is published
Free. We only write when there's a new recall that matches. Unsubscribe in one click.
Recall details
- Manufacturer
- Medtronic Neuromodulation
- Distributor
- Worldwide distribution: US (Nationwide) including Puerto Rico; and countries of: Algeria, Aruba, Austria, Belarus, Belgium, Brazil, Bulgaria, Canada, Chile, Colombia, Costa Rica, Croatia, Cyprus, Czech Republic, Denmark, Dominican Republic, Ecuador, Egypt, El Salvador, Estonia, Faroe Islands, Finland, France, French Polynesia, Georgia, Germany, Greece, Guadeloupe, Hungary, Iceland, India, Iran, Iraq, Ireland, Israel, Italy, Japan, Jordan, Kazakhstan, Kuwait, Latvia, Lebanon, Libya, Lithuania, Luxembourg, Malta, Martinique, Mexico, Morocco, Netherlands, Netherlands Antilles, New Caledonia, Norway, Oman, Pakistan, Panama, Peru, Poland, Portugal, Qatar, Reunion, Romania, Russian Federation, San Marino, Saudi Arabia, Serbia, Slovakia, Slovenia, South Africa, Spain, Sweden, Switzerland, Tunisia, Turkey, Ukraine, United Arab Emirates, United Kingdom, Uruguay, and Venezuela.
- Lot
- all serial numbers
- Risk classification
- Class II
- Area (FDA)
- device
- Country
- United States
Related recalls
- FDAInterStim SureScan MRI Leads 978A1 & 978B1: LEAD 978A128 ISTM MICRO 28CM SSMRI TINED 978A128 LEAD 978A133 ISTMOctober 28, 2026
- FDAInterStim Percutaneous Extensions EXT 3560022 ISTM II PERC 3560022 EXT 3560030 ISTM MICRO PERC 3560030October 28, 2026
- FDAMedtronic 2x4 Pocket Adaptor Kit for Deep Brain Stimulation, Model 64002; Stimulator, electrical, implanted,June 3, 2026
- FDAMedtronic 1x4 Pocket Adaptor Kit for Deep Brain Stimulation, Model 64001; Stimulator, electrical, implanted,June 3, 2026
- FDAMedtronic RestoreUltra, model 37712, and RestoreSensor, model 37714, Multi-program RechargeableNovember 20, 2013
- FDAAcetabular Pressurizer; single use and six units per box. The Acetabular Pressurizer 0206-522-000 is anNovember 20, 2013
Questions about this recall
Was Medtronic Activa PC, model 37601, and Activa RC, model 37612, Activa SC, models 37602 and 37603. Multi-program Rechargeable Neurostimulator for Deep Brain Stimulation. recalled?
Yes. FDA published recall Z-0279-2014 on November 20, 2013. Reported hazard: Medtronic notified customers regarding Loss of Stimulation of specific Implantable Neuro-Stimulators under the following specific conditions: Switching between groups with Multiple programs or creating program groups. Therapy can be restored with the Physician Recharge mode (RPM) of the Implantable Neurostimulator Recharger (INSR).
What should I do if I have this product?
Voluntary: Firm initiated · Letter