10 IN (25cm) APPX 1.1ml, SMALLBORE EXT w/3-Port NanoClave Manifold, Check Valve. 1 unit per pouch, 50 pouches per case. UDI: (01)10887709083346(17)250201(30)50(10)4718652. The Intravascular Administration Set is a single use, sterile, non-pyrogenic device which provides access for the administration of fluids from a container to a patient's vascular system through the administration set's needle or catheter (which is inserted into a vein).
Published October 28, 2020 · Recall initiated: July 6, 2020
Summary
FDA records medical device Class II recall Z-0279-2021, published October 28, 2020 and initiated July 6, 2020. Recalling firm: ICU Medical, Inc.; Status: Terminated; Reported quantity: Total of all products (Listed #1 thru 101) = 304735 units; Lot: Lot Number:4718652. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.
Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
Identification of a potential manufacturing defect on the internal surface of the NanoClave within specific lots of NanoClave sets, which may inhibit a proper seal with the NanoClave spike.
What to do
Voluntary: Firm initiated · Letter
Get an email when a new FDA recall like this is published
Free. We only write when there's a new recall that matches. Unsubscribe in one click.
Recall details
- Manufacturer
- ICU Medical, Inc.
- Distributor
- Worldwide Distribution: US (nationwide): AL, AZ, CA, CO, DE, FL,GA,ID, IL,IN, KY, LA, MA, MD, MI, MN, MO, MS, NC, NE, NH,NJ,NM,NV, NY, OH, OK, OR, PA,RI, TN, TX, UT, VA, WA, WI, and WV; and OUS countries of: Belgium, Brazil, Canada, France, Germany, Ireland, Japan, Martinique, Mexico, Poland, and Saudi Arabia.
- Lot
- Lot Number:4718652.
- Risk classification
- Class II
- Area (FDA)
- device
- Country
- United States
Related recalls
- FDAIV Sets with drip chambersSeptember 23, 2026
- FDAIV Sets with drip chambersSeptember 23, 2026
- FDAOncology sets with drip chambersSeptember 23, 2026
- FDAIV Non-Dedicated Gravity Administration Sets with drip chambersSeptember 23, 2026
- FDAMetformin Hydrochloride Extended Release 750 mg, 30 count blister cards, Rx only, Repackaged by: RemedyRepackNovember 4, 2020
- FDA226 cm (89 IN) APPX 2.7ml, MicroClave Clear, 1.2v FILTER Nanoclave T-CONN. 1 unit per pouch, 50 pouches perOctober 28, 2020
Questions about this recall
Was 10 IN (25cm) APPX 1.1ml, SMALLBORE EXT w/3-Port NanoClave Manifold, Check Valve. 1 unit per pouch, 50 pouches per case. UDI: (01)10887709083346(17)250201(30)50(10)4718652. The Intravascular Administration Set is a single use, sterile, non-pyrogenic device which provides access for the administration of fluids from a container to a patient's vascular system through the administration set's needle or catheter (which is inserted into a vein). recalled?
Yes. FDA published recall Z-0279-2021 on October 28, 2020. Reported hazard: Identification of a potential manufacturing defect on the internal surface of the NanoClave within specific lots of NanoClave sets, which may inhibit a proper seal with the NanoClave spike.
What should I do if I have this product?
Voluntary: Firm initiated · Letter