Zimmer Dental Tapered SwissPlus & SwissPlus Implant Systems surgical kit, Ref OPCST. The label shows the product is Manufactured by Zimmer Dental Inc., Carlsbad, CA. Endosseous dental implant accessories are manually powered devices intended to aid in the placement or removal of endosseous dental implants and abutments, prepare the site for placement of endosseous dental implants or abutments, aid in the fitting of endosseous dental implants or abutments, aid in the fabrication of dental prosthetics, and be used as an accessory with endosseous dental implants when tissue contact will last less than 1 hour.
Published December 27, 2017 · Recall initiated: October 11, 2017
Summary
FDA records medical device Class II recall Z-0280-2018, published December 27, 2017 and initiated October 11, 2017. Recalling firm: Zimmer Dental Inc; Status: Terminated; Reported quantity: 30 kits; Lot: Lot numbers 63666559 and 63705907. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.
Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
The device may contain a manufacturing condition affecting the geometry of the driver's hex feature which prevents it from properly engaging the fixture mount transfer and/or a dental implant.
What to do
Voluntary: Firm initiated · Letter
Get an email when a new FDA recall like this is published
Free. We only write when there's a new recall that matches. Unsubscribe in one click.
Recall details
- Manufacturer
- Zimmer Dental Inc
- Distributor
- Worldwide Distribution: US (nationwide) and countries of: Canada, Australia, Belgium, Bulgaria, Chile, Costa Rica, Egypt, France, Georgia, Germany, Hong Kong, Hungary, Iran, Ireland, Israel, Italy, Japan, Lebanon, Lithuania, Netherlands, Nicaragua, Poland, Portugal, Romania, Serbia, Shanghai, Spain, Taiwan, Thailand, Tunisia, Turkey, UK, and United Arab Emirates.
- Lot
- Lot numbers 63666559 and 63705907
- Risk classification
- Class II
- Area (FDA)
- device
- Country
- United States
Related recalls
- FDATapered Screw-Vent MTX Model TSVTB11December 27, 2017
- FDAZimmer Dental GemLock RHD2.5 Hex Driver. The label shows the product is Manufactured by Zimmer Dental Inc.,December 27, 2017
- FDASpectra System/Legacy Guided Surgery Handle Kit (Part Number: G-HK) Spectra System/Legacy Guided SurgeryDecember 27, 2017
- FDATrevo(TM) XP ProVue Procedure Pack, TREVO(TM) XP PROVUE RETRIEVER, 6 mm, 25 mm, .Excelsior(TM) XT-27(TM) 1 TIPDecember 27, 2017
- FDAGuide Sleeve 3.25 REF 2220, Guide Sleeve 3.75 REF 2221, Guide Sleeve REF 2222, Guide Sleeve REF 2223October 4, 2017
- FDAIMPLANT,1-PC,ANG,3.0X10MM, REF ZOP30A10; IMPLANT, 1-PC, ANG, 3.0X11.5MM, REF ZOP30A11; IMPLANT, 1October 4, 2017
Questions about this recall
Was Zimmer Dental Tapered SwissPlus & SwissPlus Implant Systems surgical kit, Ref OPCST. The label shows the product is Manufactured by Zimmer Dental Inc., Carlsbad, CA. Endosseous dental implant accessories are manually powered devices intended to aid in the placement or removal of endosseous dental implants and abutments, prepare the site for placement of endosseous dental implants or abutments, aid in the fitting of endosseous dental implants or abutments, aid in the fabrication of dental prosthetics, and be used as an accessory with endosseous dental implants when tissue contact will last less than 1 hour. recalled?
Yes. FDA published recall Z-0280-2018 on December 27, 2017. Reported hazard: The device may contain a manufacturing condition affecting the geometry of the driver's hex feature which prevents it from properly engaging the fixture mount transfer and/or a dental implant.
What should I do if I have this product?
Voluntary: Firm initiated · Letter