FDAFDA · Food and drugs·Recall number Z-0287-2013
PT Link Waterbath, Model#PT100. Lab Vision PT Module Plus provides an automated standarized alternative to labor intensive and time consuming dewaxing and epitote retrieval procedures.
Published November 21, 2012 · Recall initiated: May 1, 2008
Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
Dako is recalling the PT Link Waterbath because of a possible electrical hazard. Specifically, if the heater pad under the reagent tank becomes wet, it may present an electrical shock hazard when the lid is open.
What to do
Voluntary: Firm initiated · Letter
Recall details
- Manufacturer
- Dako North America Inc.
- Distributor
- Nationwide distribution: USA including states of: AK, CA, CO, CT, DC, FL, GA, HI, IA, IL IN, KS, KY, MA, MD, MI, MN, MO, NE, NJ, NY, OH, OK, OR, PA, SC, SD, TN, TX, UT, and WA.
- Lot
- Catalog #: PT-Module
- Risk classification
- Class II
- Area (FDA)
- device
- Country
- United States
Sell products like this?
Recalert watches your whole catalog for recalls like this one, automatically.