PT Link Waterbath, Model#PT100. Lab Vision PT Module Plus provides an automated standarized alternative to labor intensive and time consuming dewaxing and epitote retrieval procedures.

Published November 21, 2012 · Recall initiated: May 1, 2008

Summary

FDA records medical device Class II recall Z-0287-2013, published November 21, 2012 and initiated May 1, 2008. Recalling firm: Dako North America Inc.; Status: Terminated; Reported quantity: 136 units; Lot: Catalog #: PT-Module. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.

Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.

What's the hazard

Dako is recalling the PT Link Waterbath because of a possible electrical hazard. Specifically, if the heater pad under the reagent tank becomes wet, it may present an electrical shock hazard when the lid is open.

What to do

Voluntary: Firm initiated · Letter

Get an email when a new FDA recall like this is published

Free. We only write when there's a new recall that matches. Unsubscribe in one click.

Recall details

Manufacturer
Dako North America Inc.
Distributor
Nationwide distribution: USA including states of: AK, CA, CO, CT, DC, FL, GA, HI, IA, IL IN, KS, KY, MA, MD, MI, MN, MO, NE, NJ, NY, OH, OK, OR, PA, SC, SD, TN, TX, UT, and WA.
Lot
Catalog #: PT-Module
Risk classification
Class II
Area (FDA)
device
Country
United States

Related recalls

Questions about this recall

Was PT Link Waterbath, Model#PT100. Lab Vision PT Module Plus provides an automated standarized alternative to labor intensive and time consuming dewaxing and epitote retrieval procedures. recalled?

Yes. FDA published recall Z-0287-2013 on November 21, 2012. Reported hazard: Dako is recalling the PT Link Waterbath because of a possible electrical hazard. Specifically, if the heater pad under the reagent tank becomes wet, it may present an electrical shock hazard when the lid is open.

What should I do if I have this product?

Voluntary: Firm initiated · Letter