Cardiohelp-i, blood oxygenation and carbon dioxide removal system, Part Numbers 701072780 (Cardiohelp-i), 701068550 (Slide Rail Holder)
Published November 15, 2023 · Recall initiated: October 12, 2023
Summary
FDA records medical device Class II recall Z-0290-2024, published November 15, 2023 and initiated October 12, 2023. Recalling firm: Maquet Medical Systems USA; Status: Ongoing; Reported quantity: 1022 units; Lot: UDI-DI: 04058863074863; Serial Numbers: 90414355 through 90414868. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.
Status when published: Ongoing — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
It was found that the fixation (weld) employed by the supplier of the nut bolt to the slide rail plate was insufficient. The deficient weld allows the nut bolt to break off from the slide rail plate with slight manual force.
What to do
Voluntary: Firm initiated · Letter
Get an email when a new FDA recall like this is published
Free. We only write when there's a new recall that matches. Unsubscribe in one click.
Recall details
- Manufacturer
- Maquet Medical Systems USA
- Distributor
- US Nationwide.
- Lot
- UDI-DI: 04058863074863; Serial Numbers: 90414355 through 90414868
- Risk classification
- Class II
- Area (FDA)
- device
- Country
- United States
Related recalls
- FDACARDIOHELP-i System. Blood oxygenation and carbon dioxide removal system.May 22, 2024
- FDABEQ-HLS 7050 USA, HLS Set Advanced 7.0, Product Code/Part Number 70106.9078. Used with the CARDIOHELPFebruary 21, 2024
- FDAModel No. BEQ-HLS 5050 USA, HLS Set Advanced 5.0, Product Code/Part Number 70106.9077. Used with theFebruary 21, 2024
- FDABO-RF-32-USA RotaFlow Centrifugal Pump with SOFTLINE Coating, Product Code 701047553. For use with anFebruary 14, 2024
- FDADOP Instant Hand Sanitizer (Ethyl Alcohol 65%) with Moisturizers, Original, 4FL OZ (118 ml) bottle, Hecho enNovember 22, 2023
- FDADigitalDiagnost C90 Release 3, used to acquire, process, store, display and export digital radiographicNovember 15, 2023
Questions about this recall
Was Cardiohelp-i, blood oxygenation and carbon dioxide removal system, Part Numbers 701072780 (Cardiohelp-i), 701068550 (Slide Rail Holder) recalled?
Yes. FDA published recall Z-0290-2024 on November 15, 2023. Reported hazard: It was found that the fixation (weld) employed by the supplier of the nut bolt to the slide rail plate was insufficient. The deficient weld allows the nut bolt to break off from the slide rail plate with slight manual force.
What should I do if I have this product?
Voluntary: Firm initiated · Letter