Cardiohelp-i, blood oxygenation and carbon dioxide removal system, Part Numbers 701072780 (Cardiohelp-i), 701068550 (Slide Rail Holder)

Published November 15, 2023 · Recall initiated: October 12, 2023

Summary

FDA records medical device Class II recall Z-0290-2024, published November 15, 2023 and initiated October 12, 2023. Recalling firm: Maquet Medical Systems USA; Status: Ongoing; Reported quantity: 1022 units; Lot: UDI-DI: 04058863074863; Serial Numbers: 90414355 through 90414868. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.

Status when published: Ongoing — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.

What's the hazard

It was found that the fixation (weld) employed by the supplier of the nut bolt to the slide rail plate was insufficient. The deficient weld allows the nut bolt to break off from the slide rail plate with slight manual force.

What to do

Voluntary: Firm initiated · Letter

Get an email when a new FDA recall like this is published

Free. We only write when there's a new recall that matches. Unsubscribe in one click.

Recall details

Manufacturer
Maquet Medical Systems USA
Distributor
US Nationwide.
Lot
UDI-DI: 04058863074863; Serial Numbers: 90414355 through 90414868
Risk classification
Class II
Area (FDA)
device
Country
United States

Related recalls

Questions about this recall

Was Cardiohelp-i, blood oxygenation and carbon dioxide removal system, Part Numbers 701072780 (Cardiohelp-i), 701068550 (Slide Rail Holder) recalled?

Yes. FDA published recall Z-0290-2024 on November 15, 2023. Reported hazard: It was found that the fixation (weld) employed by the supplier of the nut bolt to the slide rail plate was insufficient. The deficient weld allows the nut bolt to break off from the slide rail plate with slight manual force.

What should I do if I have this product?

Voluntary: Firm initiated · Letter