ATS 3000 Automatic Tourniquet System w/HOSES AND LOP SENSOR Intended to be used by qualified medical professionals to temporarily occlude blood flow in a patient's extremities during surgical procedures on those extremities.
Published November 21, 2012 · Recall initiated: September 5, 2012
Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
Zimmer is conducting a voluntary removal of some Zimmer A.T.S. 3000 tourniquets after receiving complaints of both out-of-box failures and failures during surgical procedures on some of the affected units. During failure the unit will alarm, lock into a non-operational mode, and the display screen will go blank. Based on investigation, this problem appears to be due to a counterfeit chip on the
What to do
Voluntary: Firm initiated · Two or more of the following: Email, Fax, Letter, Press Release, Telephone, Visit
Recall details
- Manufacturer
- Zimmer, Inc.
- Distributor
- Worldwide Distribution - USA (nationwide) and Internationally to Belgium, Czech Republic, Egypt, France, Germany, Morocco, Netherlands, Romania, Saudi Arabia, Spain, Switzerland, Turkey, and UAE.
- Lot
- Catalog 60-3000-101-00, Serial # 3005IACJ, 3006JAEP, 3007CADR, 3007EACT, 3007EAKT, 3007GAGR, 3007HABK, 3007KAAR, 3007KAET, 3008CADE, 3008HAHC, 3008LACK, 3009AACN, 3009HABA, 3009LAEF, 3010CABH, 3010DAHG, 3010GAAC, 3010GAAJ, 3011LABL, 3011LABN, 3011LABQ, 3011LACC, 3011LACE, 3011LACF, 3011LACG, 3011LACH, 3011LACJ, 301 1LACK, 3011LACL, 3011LACM, 3011LACN, 3011LACP, 3011LACQ, 3011LACR, 3011LACS, 3011LACT, 3011LACU, 3011LACW, 3011LADA, 3011LADE, 3011LADF, 3011LADG, 3011LADH, 3011LADJ, 3011LADK, 3011LADM, 3011LADP, 3011LADQ, 3011LADR, 3011LADS, 3011LADT, 3011LADU, 3011LADW, 3011LAEA, 3011LAEB, 3011LAEC, 3011LAED, 3011LAEE, 3011LAEF, 3011LAEH, 3012AAAG, 3012AAAK, 3012AAAM, 3012AAAN, 3012AAAP, 3012AAAQ, 3012AAAR, 3012AAAS, 3012AAAT, 3012AAAU, 3012AAAW, 3012AABA, 3012AABB, 3012AABC, 3012AABE, 3012AABG, 3012AABH, 3012AABK, 3012AABL, 3012AABM, 3012AABN, 3012AABQ, 3012AABR, 3012AABS, 3012AABU, 3012AACA, 3012AACB, 3012AACC, 3012AACD, 3012AACE, 3012AACH, 3012AACJ, 3012AACM, 3012AACN, 3012AACP, 3012AACQ, 3012AACR, 3012AACS, 3012AACT, 3012AACW, 3012AADA, 3012AADB, 3012AADC, 3012AADD, 3012AADE, 3012AADF, 3012AADG, 3012BAAA, 3012BAAB, 3012BAAC, 3012BAAD, 3012BAAE, 3012BAAF, 3012BAAG, 3012BAAH, 3012BAAJ, 3012BAAK, 3012BAAL, 3012BAAM, 3012BAAN, 3012BAAP, 3012BAAQ, 3012BAAR, 3012BAAS, 3012BAAT, 3012BAAU, 3012BAAW, 3012BABA, 3012BABB, 3012BABC, 3012BABD, 3012BABE, 3012BABF, 3012BABG, 3012BABH, 3012BABJ, 3012BABK, 3012BABL, 3012BABM, 3012BABN, 3012BABP, 3012BABQ, 3012BABR, 3012BABS, 3012BABT, 3012BABU, 3012BABW, 3012BACE, 3012BACG, 3012BACH, 3012BACJ, 3012BACK, 3012BACL, 3012BACM, 3012BACN, 3012BACP, 3012BACQ, 3012BACR, 3012BACS, 3012BACT, 3012BACU, 3012BACW, 3012BADA, 3012BADB, 3012BADC, 3012BADD, 3012BADG, 3012BADH, 3012BADJ, 3012BADK, 3012BADL, 3012BADN, 3012BADP, 3012BADR, 3012BADS, 3012BADU, 3012BADW, 3012BAEA, 3012BAEB, 3012BAEC, 3012BAEH, 3012BAEJ, 3012BAEL, 3012BAEM, 3012BAEN, 3012BAEQ, 3012BAER French Catalog 60-3000 301-02 Serial # 3012 AAAE German Catalog 60-300060-301-03, Serial # 3012AAAF Spanish Catalog 60-3000-301-11, Serial # 3012AAAB, 3012AAAC, 3012AACG, 3012BACA, 3012BACB, 3012BACC, 3012BACD Dutch 60-3000-301-12, Serial # 3012AAAA, 3012AAAD, 3012EACA
- Risk classification
- Class II
- Area (FDA)
- device
- Country
- United States