ATS 3000 Automatic Tourniquet System w/HOSES AND LOP SENSOR Intended to be used by qualified medical professionals to temporarily occlude blood flow in a patient's extremities during surgical procedures on those extremities.
Published November 21, 2012 · Recall initiated: September 5, 2012
Summary
FDA records medical device Class II recall Z-0296-2013, published November 21, 2012 and initiated September 5, 2012. Recalling firm: Zimmer, Inc.; Status: Terminated; Reported quantity: 204; Lot: Catalog 60-3000-101-00, Serial # 3005IACJ, 3006JAEP, 3007CADR, 3007EACT, 3007EAKT, 3007GAGR, 3007HABK, 3007KAAR, 3007KAET, 3008CADE, 3008HAHC, 3008LACK, 3009AACN, 3009HABA, 3009LAEF, 3010CABH, 3010DAHG, 3010GAAC, 3010GAAJ, 3011LABL, 3011LABN, 3011LABQ, 3011LACC, 3011LACE, 3011LACF, 3011LACG, 3011LACH, 3011LACJ, 301 1LACK, 3011LACL, 3011LACM, 3011LACN, 3011LACP, 3011LACQ, 3011LACR, 3011LACS, 3011LACT, 3011LACU, 3011LACW, 3011LADA, 3011LADE, 3011LADF, 3011LADG, 3011LADH, 3011LADJ, 3011LADK, 3011LADM, 3011LADP, 3011LADQ, 3011LADR, 3011LADS, 3011LADT, 3011LADU, 3011LADW, 3011LAEA, 3011LAEB, 3011LAEC, 3011LAED, 3011LAEE, 3011LAEF, 3011LAEH, 3012AAAG, 3012AAAK, 3012AAAM, 3012AAAN, 3012AAAP, 3012AAAQ, 3012AAAR, 3012AAAS, 3012AAAT, 3012AAAU, 3012AAAW, 3012AABA, 3012AABB, 3012AABC, 3012AABE, 3012AABG, 3012AABH, 3012AABK, 3012AABL, 3012AABM, 3012AABN, 3012AABQ, 3012AABR, 3012AABS, 3012AABU, 3012AACA, 3012AACB, 3012AACC, 3012AACD, 3012AACE, 3012AACH, 3012AACJ, 3012AACM, 3012AACN, 3012AACP, 3012AACQ, 3012AACR, 3012AACS, 3012AACT, 3012AACW, 3012AADA, 3012AADB, 3012AADC, 3012AADD, 3012AADE, 3012AADF, 3012AADG, 3012BAAA, 3012BAAB, 3012BAAC, 3012BAAD, 3012BAAE, 3012BAAF, 3012BAAG, 3012BAAH, 3012BAAJ, 3012BAAK, 3012BAAL, 3012BAAM, 3012BAAN, 3012BAAP, 3012BAAQ, 3012BAAR, 3012BAAS, 3012BAAT, 3012BAAU, 3012BAAW, 3012BABA, 3012BABB, 3012BABC, 3012BABD, 3012BABE, 3012BABF, 3012BABG, 3012BABH, 3012BABJ, 3012BABK, 3012BABL, 3012BABM, 3012BABN, 3012BABP, 3012BABQ, 3012BABR, 3012BABS, 3012BABT, 3012BABU, 3012BABW, 3012BACE, 3012BACG, 3012BACH, 3012BACJ, 3012BACK, 3012BACL, 3012BACM, 3012BACN, 3012BACP, 3012BACQ, 3012BACR, 3012BACS, 3012BACT, 3012BACU, 3012BACW, 3012BADA, 3012BADB, 3012BADC, 3012BADD, 3012BADG, 3012BADH, 3012BADJ, 3012BADK, 3012BADL, 3012BADN, 3012BADP, 3012BADR, 3012BADS, 3012BADU, 3012BADW, 3012BAEA, 3012BAEB, 3012BAEC, 3012BAEH, 3012BAEJ, 3012BAEL, 3012BAEM, 3012BAEN, 3012BAEQ, 3012BAER French Catalog 60-3000 301-02 Serial # 3012 AAAE German Catalog 60-300060-301-03, Serial # 3012AAAF Spanish Catalog 60-3000-301-11, Serial # 3012AAAB, 3012AAAC, 3012AACG, 3012BACA, 3012BACB, 3012BACC, 3012BACD Dutch 60-3000-301-12, Serial # 3012AAAA, 3012AAAD, 3012EACA. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.
Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
Zimmer is conducting a voluntary removal of some Zimmer A.T.S. 3000 tourniquets after receiving complaints of both out-of-box failures and failures during surgical procedures on some of the affected units. During failure the unit will alarm, lock into a non-operational mode, and the display screen will go blank. Based on investigation, this problem appears to be due to a counterfeit chip on the
What to do
Voluntary: Firm initiated · Two or more of the following: Email, Fax, Letter, Press Release, Telephone, Visit
Get an email when a new FDA recall like this is published
Free. We only write when there's a new recall that matches. Unsubscribe in one click.
Recall details
- Manufacturer
- Zimmer, Inc.
- Distributor
- Worldwide Distribution - USA (nationwide) and Internationally to Belgium, Czech Republic, Egypt, France, Germany, Morocco, Netherlands, Romania, Saudi Arabia, Spain, Switzerland, Turkey, and UAE.
- Lot
- Catalog 60-3000-101-00, Serial # 3005IACJ, 3006JAEP, 3007CADR, 3007EACT, 3007EAKT, 3007GAGR, 3007HABK, 3007KAAR, 3007KAET, 3008CADE, 3008HAHC, 3008LACK, 3009AACN, 3009HABA, 3009LAEF, 3010CABH, 3010DAHG, 3010GAAC, 3010GAAJ, 3011LABL, 3011LABN, 3011LABQ, 3011LACC, 3011LACE, 3011LACF, 3011LACG, 3011LACH, 3011LACJ, 301 1LACK, 3011LACL, 3011LACM, 3011LACN, 3011LACP, 3011LACQ, 3011LACR, 3011LACS, 3011LACT, 3011LACU, 3011LACW, 3011LADA, 3011LADE, 3011LADF, 3011LADG, 3011LADH, 3011LADJ, 3011LADK, 3011LADM, 3011LADP, 3011LADQ, 3011LADR, 3011LADS, 3011LADT, 3011LADU, 3011LADW, 3011LAEA, 3011LAEB, 3011LAEC, 3011LAED, 3011LAEE, 3011LAEF, 3011LAEH, 3012AAAG, 3012AAAK, 3012AAAM, 3012AAAN, 3012AAAP, 3012AAAQ, 3012AAAR, 3012AAAS, 3012AAAT, 3012AAAU, 3012AAAW, 3012AABA, 3012AABB, 3012AABC, 3012AABE, 3012AABG, 3012AABH, 3012AABK, 3012AABL, 3012AABM, 3012AABN, 3012AABQ, 3012AABR, 3012AABS, 3012AABU, 3012AACA, 3012AACB, 3012AACC, 3012AACD, 3012AACE, 3012AACH, 3012AACJ, 3012AACM, 3012AACN, 3012AACP, 3012AACQ, 3012AACR, 3012AACS, 3012AACT, 3012AACW, 3012AADA, 3012AADB, 3012AADC, 3012AADD, 3012AADE, 3012AADF, 3012AADG, 3012BAAA, 3012BAAB, 3012BAAC, 3012BAAD, 3012BAAE, 3012BAAF, 3012BAAG, 3012BAAH, 3012BAAJ, 3012BAAK, 3012BAAL, 3012BAAM, 3012BAAN, 3012BAAP, 3012BAAQ, 3012BAAR, 3012BAAS, 3012BAAT, 3012BAAU, 3012BAAW, 3012BABA, 3012BABB, 3012BABC, 3012BABD, 3012BABE, 3012BABF, 3012BABG, 3012BABH, 3012BABJ, 3012BABK, 3012BABL, 3012BABM, 3012BABN, 3012BABP, 3012BABQ, 3012BABR, 3012BABS, 3012BABT, 3012BABU, 3012BABW, 3012BACE, 3012BACG, 3012BACH, 3012BACJ, 3012BACK, 3012BACL, 3012BACM, 3012BACN, 3012BACP, 3012BACQ, 3012BACR, 3012BACS, 3012BACT, 3012BACU, 3012BACW, 3012BADA, 3012BADB, 3012BADC, 3012BADD, 3012BADG, 3012BADH, 3012BADJ, 3012BADK, 3012BADL, 3012BADN, 3012BADP, 3012BADR, 3012BADS, 3012BADU, 3012BADW, 3012BAEA, 3012BAEB, 3012BAEC, 3012BAEH, 3012BAEJ, 3012BAEL, 3012BAEM, 3012BAEN, 3012BAEQ, 3012BAER French Catalog 60-3000 301-02 Serial # 3012 AAAE German Catalog 60-300060-301-03, Serial # 3012AAAF Spanish Catalog 60-3000-301-11, Serial # 3012AAAB, 3012AAAC, 3012AACG, 3012BACA, 3012BACB, 3012BACC, 3012BACD Dutch 60-3000-301-12, Serial # 3012AAAA, 3012AAAD, 3012EACA
- Risk classification
- Class II
- Area (FDA)
- device
- Country
- United States
Related recalls
- FDAOff-Axis Comprehensive Shoulder System Small, Augment, Off-Axis, Reamer Guide, Model/Catalog Number: 110040240July 1, 2026
- FDABrand Name: Persona Revision Product Name: Persona Revision® Trabecular Metal® Femoral Distal AugmentJuly 1, 2026
- FDAOff-Axis Alliance Glenoid, Left, 4-Peg, Augment, Reamer Guide, Model/Catalog Number: 110040444July 1, 2026
- FDAOff-Axis Comprehensive Shoulder System, Medium, Augment, Off-Axis, Reamer Guide Model/Catalog Number:July 1, 2026
- FDAHYDROCODONE BITARTRATE and ACETAMINOPHEN TABLETS, USP, 10 mg/500 mg, 100-count bottle, Rx only, Mfg. for:November 28, 2012
- FDASelect Medical Products Vinyl Powder-Free Exam Gloves, Medium Model # 6370 and Large Model # 6371, 100 perNovember 21, 2012
Questions about this recall
Was ATS 3000 Automatic Tourniquet System w/HOSES AND LOP SENSOR Intended to be used by qualified medical professionals to temporarily occlude blood flow in a patient's extremities during surgical procedures on those extremities. recalled?
Yes. FDA published recall Z-0296-2013 on November 21, 2012. Reported hazard: Zimmer is conducting a voluntary removal of some Zimmer A.T.S. 3000 tourniquets after receiving complaints of both out-of-box failures and failures during surgical procedures on some of the affected units. During failure the unit will alarm, lock into a non-operational mode, and the display screen will go blank. Based on investigation, this problem appears to be due to a counterfeit chip on the
What should I do if I have this product?
Voluntary: Firm initiated · Two or more of the following: Email, Fax, Letter, Press Release, Telephone, Visit