Millenium MPR; dual-head nuclear medicine gamma camera designed specifically for cardiac imaging
Published October 29, 2025 · Recall initiated: September 12, 2025
Summary
FDA records medical device Class II recall Z-0299-2026, published October 29, 2025 and initiated September 12, 2025. Recalling firm: GE MEDICAL SYSTEMS ISRAEL, FUNCTIONAL IMAGING; Status: Ongoing; Reported quantity: 292 units; Lot: all serial numbers in distribution. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.
Status when published: Ongoing — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
GE HealthCare has become aware that certain Nuclear Medicine systems that are past the End of Guaranteed Service could have been transported or relocated without adequate detector support. This could result in excessive stress being applied to the detector mounting mechanisms, compromising their integrity. If this occurs, it can potentially result in a detector fall and lead to life-threatening bodily injury.
What to do
Voluntary: Firm initiated · Letter
Get an email when a new FDA recall like this is published
Free. We only write when there's a new recall that matches. Unsubscribe in one click.
Recall details
- Manufacturer
- GE MEDICAL SYSTEMS ISRAEL, FUNCTIONAL IMAGING
- Distributor
- Worldwide
- Lot
- all serial numbers in distribution
- Risk classification
- Class II
- Area (FDA)
- device
- Country
- Israel
Related recalls
- FDAHADLIMA (adalimumab-bwwd) injection, 40mg/0.8 mL, 2 Single-dose Prefilled Syringes per carton, Rx only,November 5, 2025
- FDADuloxetine Delayed-Release Capsules, USP, 60 mg, a.) 90-count bottle (NDC: 51991-748-90), b.)1000-count bottleNovember 5, 2025
- FDAElscint SPX6; dual-head nuclear medicine gamma camera designed specifically for cardiac imagingOctober 29, 2025
- FDAElscint Model 4XX; dual-head nuclear medicine gamma camera designed specifically for cardiac imagingOctober 29, 2025
- FDAGE Healthcare Discovery NM/CT 670 CZT, Model/Catalog Numbers: 1) H3906AB; System, Tomography, Computed,January 22, 2025
- FDAGE Healthcare Discovery NM/CT 670 DR, Model/Catalog Numbers: 1) 5376204-70-56-2; 2) H2401AW; 3) H3100AN; 4)January 22, 2025
Questions about this recall
Was Millenium MPR; dual-head nuclear medicine gamma camera designed specifically for cardiac imaging recalled?
Yes. FDA published recall Z-0299-2026 on October 29, 2025. Reported hazard: GE HealthCare has become aware that certain Nuclear Medicine systems that are past the End of Guaranteed Service could have been transported or relocated without adequate detector support. This could result in excessive stress being applied to the detector mounting mechanisms, compromising their integrity. If this occurs, it can potentially result in a detector fall and lead to life-threatening bodily injury.
What should I do if I have this product?
Voluntary: Firm initiated · Letter