Millenium MPR; dual-head nuclear medicine gamma camera designed specifically for cardiac imaging

Published October 29, 2025 · Recall initiated: September 12, 2025

Summary

FDA records medical device Class II recall Z-0299-2026, published October 29, 2025 and initiated September 12, 2025. Recalling firm: GE MEDICAL SYSTEMS ISRAEL, FUNCTIONAL IMAGING; Status: Ongoing; Reported quantity: 292 units; Lot: all serial numbers in distribution. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.

Status when published: Ongoing — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.

What's the hazard

GE HealthCare has become aware that certain Nuclear Medicine systems that are past the End of Guaranteed Service could have been transported or relocated without adequate detector support. This could result in excessive stress being applied to the detector mounting mechanisms, compromising their integrity. If this occurs, it can potentially result in a detector fall and lead to life-threatening bodily injury.

What to do

Voluntary: Firm initiated · Letter

Get an email when a new FDA recall like this is published

Free. We only write when there's a new recall that matches. Unsubscribe in one click.

Recall details

Manufacturer
GE MEDICAL SYSTEMS ISRAEL, FUNCTIONAL IMAGING
Distributor
Worldwide
Lot
all serial numbers in distribution
Risk classification
Class II
Area (FDA)
device
Country
Israel

Related recalls

Questions about this recall

Was Millenium MPR; dual-head nuclear medicine gamma camera designed specifically for cardiac imaging recalled?

Yes. FDA published recall Z-0299-2026 on October 29, 2025. Reported hazard: GE HealthCare has become aware that certain Nuclear Medicine systems that are past the End of Guaranteed Service could have been transported or relocated without adequate detector support. This could result in excessive stress being applied to the detector mounting mechanisms, compromising their integrity. If this occurs, it can potentially result in a detector fall and lead to life-threatening bodily injury.

What should I do if I have this product?

Voluntary: Firm initiated · Letter