EMAG system Model/Catalog Number: 418591/418591U Product Description: Nucleic acid sample preparation instrument IVD, automated
Published November 4, 2026 · Recall initiated: September 11, 2026
Summary
FDA records medical device Class II recall Z-0301-2027, published November 4, 2026 and initiated September 11, 2026. Recalling firm: BioMerieux SA; Status: Ongoing; Reported quantity: 33; Lot: UDI-DI 357302BUDI000544TP Serial Numbers IM03343 IM03283 IM03104 IM03329 IM03376 IM03474 IM03321 IM03071 IM03175 IM03194 IM03515 IM03516 IM03240 IM03478 IM03479 IM03482 IM03034 IM03092 IM03378 IM03414 IM03177 IM03193 IM03506 IM3054 IM03439 IM03350 IM03380 IM03131 IM03148 IM03153 IM03275 IM03314 IM03351. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.
Status when published: Ongoing — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
Delayed or no result when analyzing clinical samples with EMAG system (ref: 418591/418591U) due to a randomly occurring software error, Error 10003 Heater timeout after updating to recent firmware SP4 Patch 2
What to do
Voluntary: Firm initiated · Letter
Get an email when a new FDA recall like this is published
Free. We only write when there's a new recall that matches. Unsubscribe in one click.
Recall details
- Manufacturer
- BioMerieux SA
- Distributor
- Worldwide - US Nationwide distribution in the states of CA, IN, KS, LA, MN, MO, NY, OH, and TX. The countries of Canada, United Kingdom, Sweden, France, Norway, Netherlands, Belgium, Denmark, Germany, Austria.
- Lot
- UDI-DI 357302BUDI000544TP Serial Numbers IM03343 IM03283 IM03104 IM03329 IM03376 IM03474 IM03321 IM03071 IM03175 IM03194 IM03515 IM03516 IM03240 IM03478 IM03479 IM03482 IM03034 IM03092 IM03378 IM03414 IM03177 IM03193 IM03506 IM3054 IM03439 IM03350 IM03380 IM03131 IM03148 IM03153 IM03275 IM03314 IM03351
- Risk classification
- Class II
- Area (FDA)
- device
- Country
- France
Related recalls
- FDAPhilips Azurion S. Model Number: 722285.November 4, 2026
- FDACareLink SmartSync Programmer, Model 24970A, Pulse generator, permanent, implantableNovember 4, 2026
- FDAMYLA software. Used to manage microbiology test workflow from the reception of requests to the transmission ofAugust 18, 2021
- FDAVIDAS FSH, Ref 30407-01, IVD, Rx. The firm name on the label is bioMerieux SA, France. For use on theJuly 4, 2018
- FDAThe VITEK® 2 Gram Positive Susceptibility Card is intended for use with the VITEK 2 Systems in clinicalJune 20, 2018
- FDAETEST OXACILLIN OX 256 US F100 In vitro diagnostic, a quantitative technique for determining the antimicrobialDecember 27, 2017
Questions about this recall
Was EMAG system Model/Catalog Number: 418591/418591U Product Description: Nucleic acid sample preparation instrument IVD, automated recalled?
Yes. FDA published recall Z-0301-2027 on November 4, 2026. Reported hazard: Delayed or no result when analyzing clinical samples with EMAG system (ref: 418591/418591U) due to a randomly occurring software error, Error 10003 Heater timeout after updating to recent firmware SP4 Patch 2
What should I do if I have this product?
Voluntary: Firm initiated · Letter