Monoject U-100 1 mL Insulin Syringe Luer-Lock with Tip Cap Soft Pack, REF 1188100777
Published November 20, 2024 · Recall initiated: September 25, 2024
Summary
FDA records medical device Class I recall Z-0310-2025, published November 20, 2024 and initiated September 25, 2024. Recalling firm: Cardinal Health 200, LLC; Status: Ongoing; Reported quantity: 371,326 units; Lot: UDI/DI 10192253034783 (EA), Lot Number 221201, exp. 2027-11-30; UDI/DI 20192253034780 (BX), Lot Number 230201, exp. 2028-01-31; UDI/DI 50192253034781 (CS), Lot Number 230202,exp. 2028-02-01. Class I means a reasonable probability of serious adverse health consequences or death.
Status when published: Ongoing — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
Certain lots of listed lots of Cardinal Health Monoject U-100 1 mL Insulin Syringe Luer-Lock with Tip Cap Soft Pack have demonstrated incompatibility with needleless Intravenous (IV) connectors. Cardinal Health recommends that they not be used for administration of IV push insulin via needleless connector.
What to do
Voluntary: Firm initiated · Letter
Get an email when a new FDA recall like this is published
Free. We only write when there's a new recall that matches. Unsubscribe in one click.
Recall details
- Manufacturer
- Cardinal Health 200, LLC
- Distributor
- Nationwide
- Lot
- UDI/DI 10192253034783 (EA), Lot Number 221201, exp. 2027-11-30; UDI/DI 20192253034780 (BX), Lot Number 230201, exp. 2028-01-31; UDI/DI 50192253034781 (CS), Lot Number 230202,exp. 2028-02-01
- Risk classification
- Class I
- Area (FDA)
- device
- Country
- United States
Related recalls
- FDACardinal Health Presource PBDS kits, Containing BD Alaris Pump Infusion Sets used with BD Alaris Pump ModuleNovember 11, 2026
- FDACardinal Health Presource PBDS kits, Containing BD Alaris Pump Infusion Sets used with BD Alaris Pump ModuleNovember 11, 2026
- FDACardinal Health Presource PBDS kits, Containing BD Alaris Pump Infusion Sets used with BD Alaris Pump ModuleNovember 11, 2026
- FDACardinal Health Presource PBDS kits, Containing BD Alaris Pump Infusion Sets used with BD Alaris Pump ModuleNovember 11, 2026
- FDABrand Name: Atlan Product Name: Atlan A350XL Model/Catalog Number: 8621600 Software Version: Not ApplicableNovember 20, 2024
- FDAmicroTargeting Insertion Tube Set, P/N 66-IT-05P, Lot 244517. The intended use is to accurately guide aNovember 20, 2024
Questions about this recall
Was Monoject U-100 1 mL Insulin Syringe Luer-Lock with Tip Cap Soft Pack, REF 1188100777 recalled?
Yes. FDA published recall Z-0310-2025 on November 20, 2024. Reported hazard: Certain lots of listed lots of Cardinal Health Monoject U-100 1 mL Insulin Syringe Luer-Lock with Tip Cap Soft Pack have demonstrated incompatibility with needleless Intravenous (IV) connectors. Cardinal Health recommends that they not be used for administration of IV push insulin via needleless connector.
What should I do if I have this product?
Voluntary: Firm initiated · Letter