BD Max Instrument, Catalog Number 441916 441927 (refurbished instrument) - Product Usage: The BD MAX System, used in conjunction with appropriate extraction kits and reagents, is capable of automated extraction and purification of nucleic acids from multiple specimen types. The BD MAX System is also capable of automated amplification and detection of target nucleic acid sequences for use in clinical, industrial, and research settings.

Published November 7, 2018 · Recall initiated: September 24, 2018

Summary

FDA records medical device Class II recall Z-0313-2019, published November 7, 2018 and initiated September 24, 2018. Recalling firm: Becton Dickinson & Co.; Status: Terminated; Reported quantity: 49; Lot: Serial Numbers: CTI1201-CTI1288. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.

Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.

What's the hazard

A trend for false positive Candida glabrata results has been confirmed. The discrepant results impact the Candida glabrata target only, when in the presence of a Trichomonas vaginalis true positive.

What to do

Voluntary: Firm initiated · Letter

Get an email when a new FDA recall like this is published

Free. We only write when there's a new recall that matches. Unsubscribe in one click.

Recall details

Manufacturer
Becton Dickinson & Co.
Distributor
Worldwide distribution - US nationwide in the states of AL, AZ, CA, CO, DC, FL, GA, IL, IN, KY, LA, MD, MI, MO, NC, NJ, NM, NY, OH, OR, PA, SC, TN, TX, UT, VA, WA, WI, and Puerto Rico. Countries of Spain, Germany, France, Austria, Denmark, Switzerland, Sweden, Italy, Kuwait, Oman, Saudi Arabia, and Turkey.
Lot
Serial Numbers: CTI1201-CTI1288
Risk classification
Class II
Area (FDA)
device
Country
United States

Related recalls

Questions about this recall

Was BD Max Instrument, Catalog Number 441916 441927 (refurbished instrument) - Product Usage: The BD MAX System, used in conjunction with appropriate extraction kits and reagents, is capable of automated extraction and purification of nucleic acids from multiple specimen types. The BD MAX System is also capable of automated amplification and detection of target nucleic acid sequences for use in clinical, industrial, and research settings. recalled?

Yes. FDA published recall Z-0313-2019 on November 7, 2018. Reported hazard: A trend for false positive Candida glabrata results has been confirmed. The discrepant results impact the Candida glabrata target only, when in the presence of a Trichomonas vaginalis true positive.

What should I do if I have this product?

Voluntary: Firm initiated · Letter