Coated VICRYL Polyglactin 910 Suture VIO 36IN(90CM) USP0(M3.5) S/A CT, Product Number J358H

Published November 13, 2024 · Recall initiated: September 25, 2024

Summary

FDA records medical device Class II recall Z-0319-2025, published November 13, 2024 and initiated September 25, 2024. Recalling firm: Ethicon, Inc.; Status: Ongoing; Reported quantity: 31,032 units; Lot: UDI-DI: 10705031237308 (each); 30705031237302 (box); Lot Number: QP2AJM. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.

Status when published: Ongoing — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.

What's the hazard

Complaints were received regarding needle pull-off and suture degradation. Analysis on returned product from this lot confirmed that some sutures were visually degraded in the package.

What to do

Voluntary: Firm initiated · Letter

Get an email when a new FDA recall like this is published

Free. We only write when there's a new recall that matches. Unsubscribe in one click.

Recall details

Manufacturer
Ethicon, Inc.
Distributor
International distribution to the country of China.
Lot
UDI-DI: 10705031237308 (each); 30705031237302 (box); Lot Number: QP2AJM
Risk classification
Class II
Area (FDA)
device
Country
United States

Related recalls

Questions about this recall

Was Coated VICRYL Polyglactin 910 Suture VIO 36IN(90CM) USP0(M3.5) S/A CT, Product Number J358H recalled?

Yes. FDA published recall Z-0319-2025 on November 13, 2024. Reported hazard: Complaints were received regarding needle pull-off and suture degradation. Analysis on returned product from this lot confirmed that some sutures were visually degraded in the package.

What should I do if I have this product?

Voluntary: Firm initiated · Letter