Surgical Planner, BrainSTORM, software version BST.9.3.0, orthopedic augmented reality
Published November 5, 2025 · Recall initiated: September 16, 2025
Summary
FDA records medical device Class II recall Z-0325-2026, published November 5, 2025 and initiated September 16, 2025. Recalling firm: Surgical Theater Inc; Status: Terminated; Reported quantity: 3 units; Lot: UDI/DI 00860003884119. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.
Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
A software anomaly, under specific conditions when large rotational alignment values are applied in 2D Image Fusion, may lead to misalignment of any secondary image layer(s) to the primary image layer.
What to do
Voluntary: Firm initiated · E-Mail
Get an email when a new FDA recall like this is published
Free. We only write when there's a new recall that matches. Unsubscribe in one click.
Recall details
- Manufacturer
- Surgical Theater Inc
- Distributor
- Worldwide - US Nationwide distribution in the states of MI, FL, DC, MN, CA, TX, NJ, PA, AZ, NY, KY, IN, WV and the countries of Italy, Germany, Israel, UK.
- Lot
- UDI/DI 00860003884119
- Risk classification
- Class II
- Area (FDA)
- device
- Country
- United States
Related recalls
- FDASuRgical Planner (SRP), software version SRP.7.12.0, imaging processing radiological systemNovember 5, 2025
- FDAAlcon PRECISION1 One-Day Contact Lenses, -2.25D.November 5, 2025
- FDAMagseed Pro 12 cm soft tissue marker. Model Number: MS2-17-1-12. Product Description: The Endomag Magseed ProNovember 5, 2025
- FDAVANTAGE GALAN 3T. CMSC Model: MRT-3020/U7. Model Number: MEXL-3020/B7.November 5, 2025
- FDABrand Name: Alcon Laboratories, Inc. Product Name: Accessory devices for Unity and Constellation surgicalNovember 5, 2025
- FDAICU Medical Chemo Lock Universal Vented Vial Spike REF: CL-70-10November 5, 2025
Questions about this recall
Was Surgical Planner, BrainSTORM, software version BST.9.3.0, orthopedic augmented reality recalled?
Yes. FDA published recall Z-0325-2026 on November 5, 2025. Reported hazard: A software anomaly, under specific conditions when large rotational alignment values are applied in 2D Image Fusion, may lead to misalignment of any secondary image layer(s) to the primary image layer.
What should I do if I have this product?
Voluntary: Firm initiated · E-Mail