ZCA All Poly Acetabular Cup Longevity Crosslinked Polyethylene, I.D. 28 mm, Neutral, O.D. Cup with Spacers, 47 mm Item Number: 00-8065-544-28
Published November 11, 2020 · Recall initiated: September 24, 2020
Summary
FDA records medical device Class II recall Z-0328-2021, published November 11, 2020 and initiated September 24, 2020. Recalling firm: Zimmer Biomet, Inc.; Status: Terminated; Reported quantity: 412 units total; Lot: All lots. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.
Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
Sterilant (hydrogen peroxide) used in acetabular cup hip prosthesis component was not evaluated to confirm the biocompatibility of the residual sterilant. Cannot rule out the potential for adverse tissue reactions.
What to do
Voluntary: Firm initiated · Letter
Get an email when a new FDA recall like this is published
Free. We only write when there's a new recall that matches. Unsubscribe in one click.
Recall details
- Manufacturer
- Zimmer Biomet, Inc.
- Distributor
- US Nationwide
- Lot
- All lots
- Risk classification
- Class II
- Area (FDA)
- device
- Country
- United States
Related recalls
- FDASpinal Rod Cutter, Item number 00392500200. The product is a nonsterile pin cutter - Product Usage: TheDecember 30, 2020
- FDAZCA All Poly Acetabular Cup Longevity Crosslinked Polyethylene, I.D. 32 mm, Snap-In O.D. Cup with Spacers, 55November 11, 2020
- FDAZCA All Poly Acetabular Cup Longevity Crosslinked Polyethylene, I.D. 32 mm, Neutral, O.D. Cup with Spacers, 55November 11, 2020
- FDAZCA All Poly Acetabular Cup Longevity Crosslinked Polyethylene, I.D. 26 mm, 10 degree inclined face, O.D. CupNovember 11, 2020
- FDAZCA All Poly Acetabular Cup Longevity Crosslinked Polyethylene, I.D. 26 mm, Neutral, O.D. Cup with Spacers, 47November 11, 2020
- FDAZCA All Poly Acetabular Cup Longevity Crosslinked Polyethylene, I.D. 22mm, Neutral, O.D. Cup with Spacers, 51November 11, 2020
Questions about this recall
Was ZCA All Poly Acetabular Cup Longevity Crosslinked Polyethylene, I.D. 28 mm, Neutral, O.D. Cup with Spacers, 47 mm Item Number: 00-8065-544-28 recalled?
Yes. FDA published recall Z-0328-2021 on November 11, 2020. Reported hazard: Sterilant (hydrogen peroxide) used in acetabular cup hip prosthesis component was not evaluated to confirm the biocompatibility of the residual sterilant. Cannot rule out the potential for adverse tissue reactions.
What should I do if I have this product?
Voluntary: Firm initiated · Letter