ALTRUS® THERMAL TISSUE FUSION, Tissue Fusion Handpiece, 5 mm O.D., 36 cm Shaft, REF 60-9522-001, STERILE -- CONMED CORPORATION 525 FRENCH RD, UTICA, NY 13502 --- The ConMed Altrus® Thermal Tissue Fusion System is comprised of a dedicated energy source and disposable handpiece used to ligate (seal) and divide (cut) blood vessels and tissue bundles that fit into the jaws of the handpiece. The Altrus® System utilizes a thermal energy platform to achieve the desired clinical effect. The ConMed Altrus® Thermal Tissue Fusion handpiece is a single use device which is provided sterile. The device uses scissors to parallel jaw closure mechanism with one flat jaw and one crowned jaw in which the vessels/tissues are grasped and through which pressure and heat are applied. It is a multi-functional device capable of vessel sealing, grasping and dissecting during open general surgical procedures. The two jaw sizes allow for applications in a variety of clinical environments for access in confined areas as well as large open areas.
Published November 27, 2013 · Recall initiated: October 23, 2013
Summary
FDA records medical device Class II recall Z-0332-2014, published November 27, 2013 and initiated October 23, 2013. Recalling firm: ConMed Corporation; Status: Terminated; Reported quantity: 147 handpieces (Domestic: 135 devices; Foreign: 12 devices); Lot: Lot 13CHB006. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.
Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
The affected lots were exposed to a higher than acceptable level of particulate matter during manufacture and subsequently distributed. ConMed evaluated the potential risk that this issue could present to the patient and concluded the risk of injury to be negligible; however exposure could cause a minor, local, self correcting infection.
What to do
Voluntary: Firm initiated · Letter
Get an email when a new FDA recall like this is published
Free. We only write when there's a new recall that matches. Unsubscribe in one click.
Recall details
- Manufacturer
- ConMed Corporation
- Distributor
- Worldwide distribution: US (Nationwide) and countries of: Canada, Germany and Italy.
- Lot
- Lot 13CHB006
- Risk classification
- Class II
- Area (FDA)
- device
- Country
- United States
Related recalls
- FDACONMED Linvatec GENESYS CrossFT Suture Anchor with Disposable Driver REF# CFBC-5502B 5.5mm GENESYs CrossFTSeptember 2, 2026
- FDAConMed AirSeal 5mm Access Port and Low Profile Obturator with Bladeless Optical Tip, 100mm Length. CatalogJune 24, 2026
- FDAConMed AirSeal 8mm Access Port and Low Profile Obturator with Bladeless Optical Tip, 100mm Length. CatalogJune 24, 2026
- FDAConMed AirSeal 12mm Access Port and Low Profile Obturator with Bladeless Optical Tip, 120mm Length. CatalogJune 24, 2026
- FDAALTRUS® THERMAL TISSUE FUSION, Tissue Fusion Handpiece, 5 mm O.D., 16 cm Shaft, REF 60-9520-001, STERILE --November 27, 2013
- FDAZeiss brand IOLMaster, Version 1-5, Biomicroscope, Slit-Lamp, AC-Powered, Product is manufactured by CarlNovember 27, 2013
Questions about this recall
Was ALTRUS® THERMAL TISSUE FUSION, Tissue Fusion Handpiece, 5 mm O.D., 36 cm Shaft, REF 60-9522-001, STERILE -- CONMED CORPORATION 525 FRENCH RD, UTICA, NY 13502 --- The ConMed Altrus® Thermal Tissue Fusion System is comprised of a dedicated energy source and disposable handpiece used to ligate (seal) and divide (cut) blood vessels and tissue bundles that fit into the jaws of the handpiece. The Altrus® System utilizes a thermal energy platform to achieve the desired clinical effect. The ConMed Altrus® Thermal Tissue Fusion handpiece is a single use device which is provided sterile. The device uses scissors to parallel jaw closure mechanism with one flat jaw and one crowned jaw in which the vessels/tissues are grasped and through which pressure and heat are applied. It is a multi-functional device capable of vessel sealing, grasping and dissecting during open general surgical procedures. The two jaw sizes allow for applications in a variety of clinical environments for access in confined areas as well as large open areas. recalled?
Yes. FDA published recall Z-0332-2014 on November 27, 2013. Reported hazard: The affected lots were exposed to a higher than acceptable level of particulate matter during manufacture and subsequently distributed. ConMed evaluated the potential risk that this issue could present to the patient and concluded the risk of injury to be negligible; however exposure could cause a minor, local, self correcting infection.
What should I do if I have this product?
Voluntary: Firm initiated · Letter