TruSystem 7000 U Operating Table (dV), Product Code 1723633
Published November 11, 2026 · Recall initiated: September 14, 2026
Summary
FDA records medical device Class II recall Z-0332-2027, published November 11, 2026 and initiated September 14, 2026. Recalling firm: Baxter Healthcare Corporation; Status: Ongoing; Reported quantity: 62 units; Lot: UDI-DI: 00887761967519; Serial Number: 110188921, 110188922, and 110206662. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.
Status when published: Ongoing — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
The main lift may unexpectedly drop from its set position to its lowest position during use. The tabletop may not be able to be raised again after the drop occurs. An uncontrolled drop of the surgical table may result in an abrupt patient displacement during surgery. Serious or critical adverse health consequences may occur, including traumatic injury, airway compromise, and surgical procedure complications, among others.
What to do
Voluntary: Firm initiated · Letter
Get an email when a new FDA recall like this is published
Free. We only write when there's a new recall that matches. Unsubscribe in one click.
Recall details
- Manufacturer
- Baxter Healthcare Corporation
- Distributor
- US Nationwide distribution in the states of IL, MO, TX.
- Lot
- UDI-DI: 00887761967519; Serial Number: 110188921, 110188922, and 110206662
- Risk classification
- Class II
- Area (FDA)
- device
- Country
- United States
Related recalls
- FDATruSystem 7500 surgical table, Model/Catalog Number: 4091000November 11, 2026
- FDAOperating Table TruSystem 7000 U, Surgical table, universal, Model/Catalog Number: 1604788November 11, 2026
- FDATruSystem 7500 Hybrid Plus (FC), Fixed-Installation Operating Table Column, Model/Catalog Number: 1854087November 11, 2026
- FDAOperating Table TruSystem 7000 U (dV), Model/Catalog Number: 1723633November 11, 2026
- FDAHalyard MINOR PACK. Model/Catalog Number: VASF105-08.November 11, 2026
- FDAMobile column TruSystem 7500 U, Product Code 1730720November 11, 2026
Questions about this recall
Was TruSystem 7000 U Operating Table (dV), Product Code 1723633 recalled?
Yes. FDA published recall Z-0332-2027 on November 11, 2026. Reported hazard: The main lift may unexpectedly drop from its set position to its lowest position during use. The tabletop may not be able to be raised again after the drop occurs. An uncontrolled drop of the surgical table may result in an abrupt patient displacement during surgery. Serious or critical adverse health consequences may occur, including traumatic injury, airway compromise, and surgical procedure complications, among others.
What should I do if I have this product?
Voluntary: Firm initiated · Letter