Back to search
FDAFDA · Food and drugs·Recall number Z-0332-2027

TruSystem 7000 U Operating Table (dV), Product Code 1723633

Published November 11, 2026 · Recall initiated: September 14, 2026

Summary

FDA records medical device Class II recall Z-0332-2027, published November 11, 2026 and initiated September 14, 2026. Recalling firm: Baxter Healthcare Corporation; Status: Ongoing; Reported quantity: 62 units; Lot: UDI-DI: 00887761967519; Serial Number: 110188921, 110188922, and 110206662. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.

Status when published: Ongoing — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.

What's the hazard

The main lift may unexpectedly drop from its set position to its lowest position during use. The tabletop may not be able to be raised again after the drop occurs. An uncontrolled drop of the surgical table may result in an abrupt patient displacement during surgery. Serious or critical adverse health consequences may occur, including traumatic injury, airway compromise, and surgical procedure complications, among others.

What to do

Voluntary: Firm initiated · Letter

Get an email when a new FDA recall like this is published

Free. We only write when there's a new recall that matches. Unsubscribe in one click.

Recall details

Manufacturer
Baxter Healthcare Corporation
Distributor
US Nationwide distribution in the states of IL, MO, TX.
Lot
UDI-DI: 00887761967519; Serial Number: 110188921, 110188922, and 110206662
Risk classification
Class II
Area (FDA)
device
Country
United States

Related recalls

Questions about this recall

Was TruSystem 7000 U Operating Table (dV), Product Code 1723633 recalled?

Yes. FDA published recall Z-0332-2027 on November 11, 2026. Reported hazard: The main lift may unexpectedly drop from its set position to its lowest position during use. The tabletop may not be able to be raised again after the drop occurs. An uncontrolled drop of the surgical table may result in an abrupt patient displacement during surgery. Serious or critical adverse health consequences may occur, including traumatic injury, airway compromise, and surgical procedure complications, among others.

What should I do if I have this product?

Voluntary: Firm initiated · Letter