Thermo SCIENTIFIC, Sensititre STP6F, Gram Negative IVD AST
Published December 14, 2022 · Recall initiated: October 20, 2022
Summary
FDA records medical device Class I recall Z-0334-2023, published December 14, 2022 and initiated October 20, 2022. Recalling firm: Remel Inc; Status: Terminated; Reported quantity: 2656 units; Lot: Lot Numbers: B2314, B2174, B2036, B1346, B1183, B1115, B1027. Class I means a reasonable probability of serious adverse health consequences or death.
Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
An internal technical investigation has confirmed that Thermo Scientific" Gram Negative IVD AST Sensititre Plate may lead to false susceptible results for Proteae Tribe (including Proteus spp., Providencia spp., Morganella Morganii) when tested with doripenem, ertapenem, meropenem, imipenem, cefepime, piperacillin/tazobactam and aztreonam.
What to do
Voluntary: Firm initiated · Letter
Get an email when a new FDA recall like this is published
Free. We only write when there's a new recall that matches. Unsubscribe in one click.
Recall details
- Manufacturer
- Remel Inc
- Distributor
- US Nationwide Distribution
- Lot
- Lot Numbers: B2314, B2174, B2036, B1346, B1183, B1115, B1027
- Risk classification
- Class I
- Area (FDA)
- device
- Country
- United States
Related recalls
- FDADewei DNA/RNA Preservation Kit, Catalog number DW-80005-1BDecember 14, 2022
- FDACarefusion V. Mueller" HARDY-STYLE BAYONET BIPOLAR FORCEPS, 0.5MM TIP, INSULATED OVERALL LENGTH 8-1/2" (215MM)December 14, 2022
- FDAThio Med w/Dex, Hem, Vit K (5ml) 100/PK Product Usage; Recommended for use in qualitative procedures as aMay 29, 2019
- FDAThio Med w/Dex, Hemin, Vit K (5ml) 20/PK Product Usage: Recommended for use in qualitative procedures as aMay 29, 2019
- FDABlood Agar (TSA w/ 5% Sheep Blood) Plate, For In Vitro Diagnostic Use, Remel, Lenexa, KS 66215 Product Usage:May 3, 2017
- FDARemel, Haemophilus Test Medium (Agar) in A) 100 mm REF R01503, and B) 150 mm plates REF R04033,March 15, 2017
Questions about this recall
Was Thermo SCIENTIFIC, Sensititre STP6F, Gram Negative IVD AST recalled?
Yes. FDA published recall Z-0334-2023 on December 14, 2022. Reported hazard: An internal technical investigation has confirmed that Thermo Scientific" Gram Negative IVD AST Sensititre Plate may lead to false susceptible results for Proteae Tribe (including Proteus spp., Providencia spp., Morganella Morganii) when tested with doripenem, ertapenem, meropenem, imipenem, cefepime, piperacillin/tazobactam and aztreonam.
What should I do if I have this product?
Voluntary: Firm initiated · Letter